Performance evaluation of a portable bioimpedance cardiac output monitor for measuring hemodynamic changes in athletes during a head-up tilt test

Author:

Cheung Cara H. Y.1,Khaw May L.2,Tam Victor C. W.1,Ying Michael T. C.1,Lee Shara W. Y.1ORCID

Affiliation:

1. Department of Health Technology and Informatics, Faculty of Health and Social Sciences, The Hong Kong Polytechnic University, Hong Kong Special Administrative Region, China

2. Tasmanian School of Medicine, University of Tasmania, Hobart, Tasmania

Abstract

The use of impedance cardiography to monitor physiological changes in sports is rarely reported. Using head-up tilt test, we evaluated a portable noninvasive impedance cardiography device (PhysioFlow) by comparing it with a reference Doppler monitor (USCOM). Accuracy in tracking hemodynamic changes deteriorated with higher tilt, implying a gravitational influence on its performance. Stroke volume measurements were overestimated, but the changes were underestimated. Despite its convenient physical features, the suitability of PhysioFlow for sports use is questionable.