Mimicking a flow-limited human upper airway using a collapsible tube: relationships between flow patterns and pressures in a respiratory model

Author:

Zhu Kaixian1ORCID,Farré Ramon23,Katz Ira45,Hardy Sébastien1,Escourrou Pierre6

Affiliation:

1. Centre Explor, Air Liquide Healthcare, Gentilly, France

2. Unit of Biophysics and Bioengineering, Faculty of Medicine and Health Sciences, University of Barcelona-Institut Investigacions Biomediques August Pi Sunyer, Barcelona, Spain

3. Centro de Investigacion Biomedica en Red (CIBER) de Enfermedades Respiratorias, Madrid, Spain

4. Medical Research & Development, Air Liquide Santé International, Centre de Recherche Paris-Saclay, Les Loges-en-Josas, France

5. Department of Mechanical Engineering, Lafayette College, Easton, Pennsylvania

6. Department of Physiology, Université Paris-Saclay, AP-HP Hôpital Antoine-Béclère, Clamart, France

Abstract

The upper airway (UA) in humans is commonly modeled as a Starling resistor. However, negative effort dependence (NED) observed in some patients with obstructive sleep apnea (OSA) contradicts predictions based on the Starling resistor model in which inspiratory flow is independent of inspiratory driving pressure when flow is limited. In a respiratory bench model consisting of a collapsible tube and an active lung model (ASL5000), inspiratory flow characteristics were investigated in relation to upstream, downstream, and extra-luminal pressures (denoted as Pus, Pds, and Pout, respectively) by varying inspiratory effort (muscle pressure) from −1 to −20 cmH2O in the active lung. Pus was provided by a constant airway pressure device and varied from 4 to 20 cmH2O, and Pout was set at 10 and 15 cmH2O. Upstream resistance at onset of flow limitation and critical transmural pressure (Ptm) corresponding to opening of the UA were found to be independent of Pus, Pds, and Pout. With fixed Ptm, when Pds fell below a specific value (Pds′), inspiratory peak flow became constant and independent of Pds. NED plateau flow patterns at mid-inspiration (V̇n) were produced within the current bench setting when Pds fell below Pds′. V̇n was proportional to Pds, and the slope (ΔV̇n/ΔPds) increased linearly with Ptm. Ptm and Pds were the two final independent determinants of inspiratory flow. Our bench model closely mimics a flow-limited human UA, and the findings have implications for OSA treatment and research, especially for bench-testing auto-titrating devices in a more physiological way. NEW & NOTEWORTHY A respiratory model consisting of a collapsible tube was used to mimic a flow-limited human upper airway. Flow-limited breathing patterns including negative effort dependence were produced. Transmural and downstream pressures acting on the tube are the two independent determinants of the resulting inspiratory flow during flow limitation. The findings have implications for obstructive sleep apnea treatment and research, especially for bench-testing auto-titrating devices in a more physiological way.

Publisher

American Physiological Society

Subject

Physiology (medical),Physiology