Dye-dilution cardiac output without blood withdrawal in the conscious rabbit
-
Published:1978-08-01
Issue:2
Volume:235
Page:H258-H261
-
ISSN:0363-6135
-
Container-title:American Journal of Physiology-Heart and Circulatory Physiology
-
language:en
-
Short-container-title:American Journal of Physiology-Heart and Circulatory Physiology
Author:
Boerboom L. E.,Boelkins J. N.
Abstract
Cardiac output can be measured in the conscious rabbit without blood withdrawal by use of a modified dye-dilution method. Blood flows continuously from the central ear artery through a Silastic catheter that passes through a modified cuvette lumen and back into an ear vein. Blood flow through this shunt was 1.5 ml/min. This modified technique was evaluated by comparison with a simultaneously performed cardiac output measurement using radioactive microspheres and by making consecutive dye-dilution measurements at short time intervals. A correlation coefficient of 0.96 was obtained when comparing dye-dilution and microsphere cardiac outputs. The difference between 11 paired measurements was 7.09 +/- 5.24% (mean +/- SD). Consecutive outputs measured within 3 min of each other by the flow-through technique had a correlation coefficient of 0.99 and a difference between the first and second determinations of 3.52 +/- 2.38%. The method permits dye-dilution cardiac output measurement in the conscious rabbit while avoiding the potential error caused by blood withdrawal from small animals.
Publisher
American Physiological Society
Subject
Physiology (medical),Cardiology and Cardiovascular Medicine,Physiology
Cited by
3 articles.