Validity of thermal dilution technique for measurement of cardiac output in rats

Author:

Kissling G.1,Ross C.1,Brandle M.1

Affiliation:

1. Physiologisches Institut II, Universitat Tubingen, Germany.

Abstract

The validity of the thermal dilution technique for the measurement of cardiac output was verified in experiments on a circulation model and on anesthetized rats under open- and closed-chest conditions. In the circulation model thermal dilution was compared with direct (Fdir) and electromagnetic (F(elm)) flowmetry. Flow values measured in the circulation model with the thermal dilution (Fth) technique correspond well with direct flowmetry (Fth = 0.92 Fdir + 7.0; r = 0.888) and with electromagnetic flowmetry (Fth = 0.95 F(elm) + 1.2; r = 0.990). In the anesthetized rat cardiac output was determined with thermal dilution and simultaneously with Fick's method and/or with electromagnetic flowmetry. Fick's method and electromagnetic flowmetry resulted in identical cardiac output values (COFick = 0.95 COelm; r = 0.865), whereas the thermal dilution technique yielded unequivocally higher values. The extent of overestimation is much more pronounced at low cardiac output than at a high output. The study clearly demonstrates that this overestimation is due to heat diffusion, which is obviously of greater significance in small animals than in large animals or humans. Therefore, the thermal dilution technique is not appropriate for the measurement of cardiac output in the rat.

Publisher

American Physiological Society

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine,Physiology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3