Affiliation:
1. Department of Physiology, Charing Cross and Westminster MedicalSchool, London, United Kingdom.
Abstract
Conduction velocity and recovery of excitability are central facets of reentry arrhythmias, and yet there are no satisfactory techniques for the simultaneous measurement of both from the same area of myocardium. We have developed an electrode arrangement that allows the simultaneous recording of conduction velocity, repolarization of the myocardium together with an index of dispersion, and direction of activation of the myocardium. Three silver/silver chloride electrodes were arranged in an equilateral triangle with a reference electrode at the center. From this arrangement three monophasic action potentials were recorded. From the time of arrival of the wavefront of activation at each electrode the direction of activation and conduction velocity were calculated in real time by a computer. There was a good correlation for the in vivo signals from the circular electrode and the new electrode both for conduction velocity (r = 0.99, P < 0.001) and for direction of activation (r = 0.99, P < 0.001). This new mathematical method and electrode design allows the simultaneous measurement of conduction velocity and direction and monophasic action potential, and this can give a beat-by-beat indication of wavelength and dispersion of action potential duration.
Publisher
American Physiological Society
Subject
Physiology (medical),Cardiology and Cardiovascular Medicine,Physiology
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献