Mis-sizing of stent promotes intimal hyperplasia: impact of endothelial shear and intramural stress

Author:

Chen Henry Y.123,Sinha Anjan K.4,Choy Jenny S.23,Zheng Hai23,Sturek Michael52,Bigelow Brian6,Bhatt Deepak L.7,Kassab Ghassan S.14523

Affiliation:

1. Weldon School of Biomedical Engineering, Purdue University, West Lafayette;

2. Cellular and Integrative Physiology, Indiana University School of Medicine, Indianapolis;

3. Department of Biomedical Engineering, Indiana University-Purdue University Indianapolis, Indianapolis, Indiana

4. Departments of 2Medicine (Cardiology),

5. Surgery, and

6. Heart Center, St. Vincent Hospital, Indianapolis, Indiana;

7. Veterans Affairs Boston Healthcare System, Brigham and Women's Hospital, and Harvard Medical School, Boston, Massachusetts; and

Abstract

Stent can cause flow disturbances on the endothelium and compliance mismatch and increased stress on the vessel wall. These effects can cause low wall shear stress (WSS), high wall shear stress gradient (WSSG), oscillatory shear index (OSI), and circumferential wall stress (CWS), which may promote neointimal hyperplasia (IH). The hypothesis is that stent-induced abnormal fluid and solid mechanics contribute to IH. To vary the range of WSS, WSSG, OSI, and CWS, we intentionally mismatched the size of stents to that of the vessel lumen. Stents were implanted in coronary arteries of 10 swine. Intravascular ultrasound (IVUS) was used to size the coronary arteries and stents. After 4 wk of stent implantation, IVUS was performed again to determine the extent of IH. In conjunction, computational models of actual stents, the artery, and non-Newtonian blood were created in a computer simulation to yield the distribution of WSS, WSSG, OSI, and CWS in the stented vessel wall. An inverse relation ( R2 = 0.59, P < 0.005) between WSS and IH was found based on a linear regression analysis. Linear relations between WSSG, OSI, and IH were observed ( R2 = 0.48 and 0.50, respectively, P < 0.005). A linear relation ( R2 = 0.58, P < 0.005) between CWS and IH was also found. More statistically significant linear relations between the ratio of CWS to WSS (CWS/WSS), the products CWS × WSSG and CWS × OSI, and IH were observed ( R2 = 0.67, 0.54, and 0.56, respectively, P < 0.005), suggesting that both fluid and solid mechanics influence the extent of IH. Stents create endothelial flow disturbances and intramural wall stress concentrations, which correlate with the extent of IH formation, and these effects were exaggerated with mismatch of stent/vessel size. These findings reveal the importance of reliable vessel and stent sizing to improve the mechanics on the vessel wall and minimize IH.

Publisher

American Physiological Society

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine,Physiology

Cited by 53 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献