Daily patient geometry correction: application of NAL and eNAL protocols

Author:

Dąbrowski Andrzej12,Zielińska-Dąbrowska Sylwia2,Kuszewski Tomasz12,Lis Krzysztof2

Affiliation:

1. Physics Instytute , Jan Kochanowski University , ul. Uniwersytecka 7, 25-406 Kielce , Poland

2. Medical Physics Department , Holycross Cancer Centre , Artwińskiego 3, 25-374 Kielce , Poland

Abstract

Abstract Purpose: To test the NAL and eNAL correction protocols using daily patient setup displacements. Methods and material: In total, the analysis was performed for 749 and 797 kV CBCT images for gynecological and prostate patients, respectively, each of 30 patients. After the planning procedure, patients were set up on the treatment table in the treatment position every day. The on-line correction protocol was applied. KV CBCT images were acquired by means of x-ray lamp mounted orthogonally on Linac. Patient setup displacement was assigned. NAL and eNAL corrections protocols were simulated using daily data from online corrections for these two groups of patients. The overall systematic error and random error were calculated for each direction. Results: For the prostate group, the random errors for daily Raw data (no correction) in LAT, LONG, and VERT directions were 2.0 mm, 1.6 mm, and 3.2 mm, respectively. For NAL and eNAL protocols, they were in the range of 1.8 mm to 3.2 mm. For the gynecological group, the random errors were: for daily Raw data 2.2 mm, 1.7 mm, and 3.2 mm, respectively. For NAL and eNAL protocols, they were in the range of 2.0 to 3.4 mm. For the prostate group, values of systematic errors 1.8 mm, 1.8 mm, and 3.3 mm, respectively for Raw data. For NAL and eNAL protocols, these values were less than 1.8 mm. For the gynecological group, the systematic errors were 2.6 mm, 2.3 mm, and 2.8 mm, respectively, for Raw data. For NAL ana eNAL protocols less than 1.8 mm. For the gynecological group, for Raw data, 45% of the total displacement vectors exceeded 5 mm, whereas only 25% did after the NAL procedure and 29% after the eNAL procedure. For the prostate group, for Raw data, 34% of the total displacement vectors exceeded 5 mm, whereas only 22% did after NAL procedure and 28% after eNAL procedure Conclusions: For gynecological and prostate cancer patients, the NAL and eNAL correction protocols can be safely applied to substantially reduce setup errors.

Publisher

Walter de Gruyter GmbH

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3