Effect of the Small Field of View and Imaging Parameters to Image Quality and Dose Calculation in Adaptive Radiotherapy

Author:

Ubaidillah Allam1,Fauzia Annisa Rahma2,Purnomo Adi Teguh3,Nasution Nuruddin2,Wibowo Wahyu Edy2,Pawiro Supriyanto Ardjo1ORCID

Affiliation:

1. 1 Department of Physics, Faculty of Mathematics and Natural Sciences, Universitas Indonesia , Depok , West Java , , Indonesia

2. 2 Department of Radiation Oncology, Faculty of Medicine, Universitas Indonesia, dr. Cipto Mangunkusumo General Hospital , Jakarta , , Indonesia

3. 3 PT Elekta Medical Solution , Cowell Tower 7th Floor Unit 2A , Jakarta , , Indonesia

Abstract

Abstract Background The use of cone beam computed tomography (CBCT) for dose calculation in adaptive radiotherapy has been investigated in many studies. Proper acquisition and reconstruction of preset parameters could improve the accuracy of dose calculation based on CBCT images. This study evaluated the impact of the modified image acquisition and preset reconstruction parameter available in X-Ray Volumetric Imaging (XVI) to improve CBCT image quality and dose calculation accuracy. Materials and methods Calibration curves were generated by scanning the CIRS phantom using CBCT XVI Elekta 5.0.4 and Computed Tomography (CT) Simulator Somatom, which served as CT image reference. Rando and Catphan 503 phantoms were scanned with various acquisition and reconstruction parameters for dose accuracy and image quality tests. The image quality test is uniformity, low contrast visibility, spatial resolution, and geometrical scale test for each image by following the XVI image quality test module. Results Acquisition and reconstruction parameters have an impact on the Hounsfield Unit (HU) value that is used as the HU-Relative Electron Density (RED) calibration curve. The dose difference for all the calibration curves was within 1% and passed the gamma passing rate. Images acquired using 120 kVp, F1 (with Bowtie Filter), and 50 mA (F1-120-50-10) scored the highest Gamma Index (GI) of 98.5%. For the image quality test, it scored 1.20% on the uniformity test, 2.14% on the low contrast visibility test, and 11 lp/cm on the spatial resolution test. However, F1-120-50-10 reconstructed with different reconstructions scored 3.83% and 4 lp/cm in contrast and spatial resolution test, respectively. Conclusion CBCT reconstruction parameters work as a scatter correction. It could improve the dose accuracy and image quality. Nevertheless, without adequate CBCT acquisition protocols, it would produce an image with high uncertainty and cannot be fixed with reconstruction protocols. The F1-120-50-10 protocols generate the highest dose accuracy and image quality.

Publisher

Walter de Gruyter GmbH

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3