Evaluation of dosimetric implications of Pareto and constrained mode of optimization for Monaco TPS generated VMAT plans in post operated carcinoma of the left breast

Author:

Srivastava Anoop Kumar1,Bharati Avinav1,Rastogi Madhup1,Mishra Surendra Prasad1,Khurana Rohini1,Hadi Rahat1,Gandhi Ajeet Kumar1,Mishra Lalatendu1

Affiliation:

1. Department of Radiation Oncology , Dr. Ram Manohar Lohia Institute of Medical Sciences , Lucknow India -226010

Abstract

Abstract Intensity-modulated radiotherapy (IMRT) is being practiced for the last several years with a special approach for radiation therapy in post-mastectomy breast cancer patients. Meeting the cardiac dose constraints has always been a challenge during radiotherapy planning by both IMRT and VMAT (volumetric modulated arc therapy) of post-mastectomy left breast patients. With the advancement in IMRT planning techniques, it has been modified to VMAT with more degrees of freedom for modulation and is being utilised more frequently. This helps in obtaining a suitable plan for achieving both the dose homogeneity in target volume and dose constraints to Organ at Risk (OAR). 10 Patients with carcinoma of the left breast (post-mastectomy) were selected for this study. VMAT treatment plans for these patients were generated for 6 MV photons on the Monaco treatment planning system (TPS) using two types of optimization modes i.e. Pareto and Constrained mode available in Monaco TPS. For comparative dosimetric evaluation of the efficacy of these two types of optimization modes similar calculation algorithms, calculation grids, arcs, and beam sequencing parameters were used for generating treatment plans. The dosimetric quantities such as volume receiving more than 95% of the prescribed dose (V95), volume receiving more than 107% of the prescribed dose (V107) and Maximum dose (Dmax) for target volume, mean dose (Dmean) for heart, volume receiving more than 30 Gy (V30) volume receiving more than 20 Gy (V20) volume receiving more than 5 Gy (V5) for ipsilateral lung and total monitor units delivered were analysed for both optimization modes. A judicious mix of multiple planning parameters and variables using these two modes of optimization was applied and recorded. Both optimization modes yielded similar outcomes. However, Pareto mode has shown better coverage for planning target volume (PTV) with comparable doses to OARs.

Publisher

Walter de Gruyter GmbH

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3