Ultraviolet ray irradiation duration for the pre-exposure of Gafchromic EBT2

Author:

Katsuda Toshizo1,Gotanda Rumi2,Gotanda Tatsuhiro2,Akagawa Takuya3,Tanki Nobuyoshi1,Kuwano Tadao4,Noguchi Atsushi1,Yabunaka Kouichi5

Affiliation:

1. Faculty of Health Sciences , Butsuryo College of Osaka , 3-33, Otorikita-machi Nishi-ku, Sakai-city , Osaka , 593-8328 , Japan

2. Faculty of Health Science and Technology , Kawasaki University of Medical Welfare , Okayama , Japan

3. Department of Radiological Technology , Tokushima Red Cross Hospital , 103, Irinokuchi, Komatsujshima, Tokushima , 773-8502 , Japan

4. Graduate School of Health Sciences , Okayama University , 2-5-1, Shikata-cyo, Kita-ku, Okayama-city , Okayama , 700-8558 , Japan

5. Graduate School of Medicine , the University of Tokyo , 7-3-1 Hongo, Bunkyou-Ku, Tokyo , 113-0033 , Japan

Abstract

Abstract The strength and density change of the ultraviolet (UV) ray of Gafchromic EBT2 were investigated. Previous studies suggested that UV-A rays can be substituted for the x-ray double-exposure technique to correct Gafchromic EBT2’s non-uniformity error. In this study, we aimed to determine the appropriate strength of UV-A rays for irradiating an active layer that would correct the non-uniformity error of Gafchromic EBT2. UV-A rays with a wavelength of 375 nm were used to irradiate Gafchromic EBT2 in various durations, and the resulting density change was investigated. To correct Gafchromic EBT2’s non-uniformity error, a pre-irradiation with a UV-A lamp was conducted at a distance of 72 cm for 30 min. To determine the most appropriate irradiation duration, a UV light-emitting diode generating UV-A of 375 nm was used to irradiate the Gafchromic EBT2 film with varying durations of 1, 2, 3, 4, 5, 10, 15, 20, 25, and 30 min at a distance of 5.3 cm. A 12.7 diameter region of interest was set by the irradiation area, and a histogram of pixel values was created. The condition options were decided based on two important requirements: 1) no zero values of the mode and seconds exist, and 2) the 1/10 value of the mode intersects both histogram sleeves. In the case of Gafchromic EBT2, the irradiation strength was 85.43 mJ/cm2 for one minute in which the pixel value of mean ± SD was 255.34 ± 213.29. The irradiation duration of 4 min was the border duration of the above two conditions. When a UV ray of 375 nm wavelength is used to irradiate Gafchromic EBT2 as a substitute for x-ray exposure, the 4-min pre-irradiation duration (341.74 mJ/cm2) is demonstrably sufficient.

Publisher

Walter de Gruyter GmbH

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3