Affiliation:
1. Department of Radiation Sciences , Yasuj University of Medical Sciences , Yasuj , Iran
Abstract
Abstract
Introduction: Although in many developed countries, Analog radiography (AR) is replaced with digital radiography (DR) but AR is still widely used in many countries included Iran. Therefore, dosimetrically assessment of delivered dose is very important to avoid unnecessary patient dose.
Materials and Methods: In this study, all imaging centers in Kohgiluyeh and Boyer-Ahmad were selected. The initial information included the mean kVp and mAs used by the personnel to perform each radiological procedure were gathered through a questionnaire. Barracuda dosimeter was then used to measure Incident air kerma (ki). Data obtained from digital radiography (DR) and analogue radiography (AR) were then analyzed and compared to each other.
Results: The mean incident air kerma (ki) for five radiological procedures (chest AP&Lat, Skull AP&Lat, Lumbar spine AP&Lat, Thoracic spine AP&Lat and Pelvis) in digital devices were 0.38&1.34, 2.1&1.94, 4.99&7.83, 4.18& 6.41 and 4.33 mGy and those for analogue devices were 0.7&1.28, 3.05&3.02, 7.25&9.9, 7.125&8.36 and 5.36 mGy, respectively.
Discussion and Conclusion: The use of low kVp or high mAs is one of the reasons to increase the incident air kerma (ki) in analogue methods comparing to digital methods in all procedures except the chest (in Lateral view). Also the results, surprisingly, showed that in some of the analogue methods incident air kerma (ki) was less than digital methods which is most probably because of the auto-exposure conditions.
Reference26 articles.
1. [1] Huda W, Nickoloff EL, Boone JM. Overview of patient dosimetry in diagnostic radiology in the USA for the past 50 years. Med Phys. 2008;35(12): 5713-5728.10.1118/1.3013604
2. [2] Bahreyni Toossi MT, Esmaili S. Estimation of entrance surface doses (ESD) for common medical X-ray diagnostic examinations in radiological departments in Mashhad-Iran. In Shielding aspects of accelerators, targets and irradiation facilities-Satif-10. Nuclear Energy Agency of the OECD (NEA): Organisation for Economic Co-Operation and Development - Nuclear Energy Agency. 2010.
3. [3] Bouzarjomehri F, Dashti M, Zare M. Radiation exposure of the Yazd population from medical conventional X-ray examinations. Iran J Radiat Res. 2007;4(4):195-200.
4. [4] Bouzarjomehri F. Patient dose in routine X-ray examinations in Yazd state. Iran J Radiat Res. 2004;1(4):199-204.
5. [5] Uffmann M, Schaefer-Prokop C. Digital radiography: the balance between image quality and required radiation dose. EurJ Radiol. 2009;72(2): 202-208.
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献