Multicenter Comparison of Seven 25Oh Vitamin D Automated Immunoassays / Multicentrično Poređenje Sedam Automatizovanih Imunoeseja Za 25Oh Vitamin D

Author:

Lippi Giuseppe,Salvagno Gian Luca,Fortunato Antonio,Dipalo Mariella,Aloe Rosalia,Rin Giorgio Da,Giavarina Davide

Abstract

SummaryBackground: The measurement of 25OH vitamin D continues to grow in clinical laboratories. The aim of this multicenter study was to compare the results of seven automated commercial immunoassays with a reference HPLC technique.Methods: One hundred and twenty consecutive outpatient serum samples were centrifuged, divided in aliquots, frozen and shipped to the participating laboratories. 25OH Vita - min D was measured with a reference HPLC system and with seven automated commercial immunoassays (Roche Cobas E601, Beckman Coulter Unicel DXI 800, Ortho Vitros ES, DiaSorin Liaison, Siemens Advia Centaur, Abbott Architect i System and IDS iSYS).Results: Compared to the reference method, the regression coefficients ranged from 0.923 to 0.961 (all p<0.001). The slope of Deming fit ranged from 0.95 to 1.06, whereas the intercept was comprised between -15.2 and 9.2 nmol/L. The bias from the reference HPLC technique varied from 14.5 to 8.7 nmol/L. The minimum performance goal for bias was slightly exceeded by only one immunoassay. The agreement between HPLC and the different immunoassays at 50 nmol/L 25OH Vitamin D varied between 0.61 and 0.85 (all p<0.001). The percentage of samples below this cut-off was significantly different with only one immunoassay.Conclusions:The excellent correlation with the reference HPLC technique attests that all seven automated immuno - assays may be reliably used for routine assessment of 25OH-D in clinical laboratories. The significant bias among the different methods seems mostly attributable to the lack of standardization and calls for additional efforts for improving harmonization of 25OH-D immunoassays.

Publisher

Centre for Evaluation in Education and Science (CEON/CEES)

Reference19 articles.

1. Comparison of three different methods for vitamin determination and vitamin D status in general population;Jovičić;J Med Biochem,2012

2. thrombosis and hemostasis : more than skin deep;Targher;Semin Thromb Hemost,2012

3. Performance evaluation of Siemens ADVIA Centaur and Roche MODULAR Analytics Total OH Vitamin D assays;Chen;Clin Biochem,2012

4. Four automated OH total vitamin D immunoassays and commercial liquid chromatography tandem - mass spectrometry in Finnish population;Koivula;Clin Lab,2013

5. An assessment of hydroxy vitamin D measurements in comparability studies conducted by the Vitamin D Metabolites Quality Assurance Program;Bedner;Clin Chim Acta,2013

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献