Abstract
Abstract
The article examines: (i) the reasons of error due to thermoelectric inhomogeneity of electrodes of thermocouples acquired during prolonged use; (ii) the neural network method of error correction based on a generalization of verification results in several temperature fields; (iii) the method of investigating the impact of changing the speed of the conversion characteristic drift of thermocouple on error correction; (iv) results of this investigation. It is shown that residual error for type K thermocouples at the 5 % level of significance does not exceed μ±0.46 oС and one at the 10 % level of significance does not exceed ±0.25 °С
Subject
Instrumentation,Biomedical Engineering,Control and Systems Engineering
Cited by
35 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献