Anti-thyroid peroxidase (TPO) antibodies – Comparative analysis of two automatic methods, ECLIA and CMIA

Author:

Manescu Ion Bogdan12,Luca Andreea3,Hutanu Adina12,Truta Andreea3,Dobreanu Minodora12

Affiliation:

1. 1 Department of Laboratory Medicine , George Emil Palade University of Medicine, Pharmacy, Science, and Technology of Targu Mures , Romania

2. 2 Department of Immunology , Center for Advanced Medical and Pharmaceutical Research , Romania

3. 3 Department of Laboratory Medicine , County Emergency Clinical Hospital of Targu Mures , Romania

Abstract

Abstract Introduction: Anti-thyroid peroxidase autoantibodies (TPO) is an essential diagnostic tool for autoimmune disorders of the thyroid gland. However, TPO results are not always comparable due to differences between methods. Here, we aimed to investigate the differences between two modern laboratory methods for TPO measurement: electrochemiluminescence (ECLIA) and chemiluminescence microparticle (CMIA) immunoassays. Methods: A total of 234 serum samples were tested on two methods: Cobas-e601 (ECLIA) and Alinity i (CMIA). TPO results were compared statistically both quantitatively and qualitatively (results were coded as positive/negative, according to ECLIA/CMIA reference ranges. Results: The precisions of both methods were acceptable compared with the claims of the manufacturer. There was a very strong, but unsatisfactory correlation between the two methods (Pearson r=0.85). Passing-Bablok regression revealed a significant deviation from linearity (Cusum p<0.01) and an unacceptable quantitative relationship: intercept −7.61, slope 1.10. Moreover, a visual analysis of overall and medical decision level-focused Bland-Altman plots confirmed the lack of quantitative agreement. As for the qualitative analysis, the concordance rate between methods was 218/234 (93.1%). The agreement was considered good to very good according to the inter-rater agreement test: weighted Cohen κ = 0.805. Conclusions: The qualitative agreement between Cobas-e601 (ECLIA) and Alinity i (CMIA) was good, therefore the two methods may be used indiscriminately for initial testing of patients suspected of thyroid gland autoimmune diseases. However, due to poor quantitative agreement, the two methods should not be used interchangeably for monitoring as the results may mislead both physicians and patients, possibly leading to medical errors.

Publisher

Walter de Gruyter GmbH

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3