Affiliation:
1. 1Aristotle University of Thessaloniki
Abstract
AbstractAn HPLC method for the quality control of famotidine (FMT) containing formulations has been developed and validated. The combination of a short monolithic column (Chromolith® RP-18e 50 × 4.6 mm i.d.) and an elevated flow rate (3.0 mL min−1) enabled the proposal of a high-throughput analytical scheme capable of reliable operation in a demanding industrial environment. Detection was carried out at 265 nm. Thorough validation of the method included linearity (5–150%), limits of detection (0.13%) and quantification (0.41%), selectivity, precision (within- and day-to-day), accuracy and ruggedness. The new method was applied successfully to the analysis of samples (assay, dissolution, dosage & blending uniformity) during the production of four validation batches of FMT-containing tablets.
Subject
Materials Chemistry,General Chemistry
Reference28 articles.
1. T. J. Humphries, Scand. J. Gastroenterol. 22, 55 (1987)
2. B. Simon, P. Muller, H.G. Dammann, Innere Medizin 11, 81 (1984)
3. T.J. Humphries and G.J. Merritt, Aliment. Pharmacol. Ther. 13, 18 (1999)
4. P.D. Tzanavaras, D.G. Themelis, Anal. Chim. Acta 588, 1 (2007)
5. D. Guillarme, D.T.T. Nguyen, S. Rudaz, J-L. Veuthey, Eur. J. Pharm. Biopharm. 68, 430 (2008)
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献