The accuracy (trueness and precision) of Bellus3DARC-7 and an in-vivo analysis of intra and inter-examiner reliability of digital and manual anthropometry

Author:

Doan Jenny1,Finlay Nessa1,Lee Amy1,Ethell Thomas1,Gandedkar Narayan1,Darendeliler Mehmet Ali1

Affiliation:

1. Discipline of Orthodontics Dental School, Faculty of Medicine and Health, University of Sydney , Sydney Dental Hospital, Sydney Local Health District, Surry Hills, NSW , Sydney , Australia

Abstract

Abstract Background Three dimensional (3D) facial imaging is becoming a popular method of facial analysis and a means of anthropometry. There is potential for 3D imaging to provide enough diagnostic information to parallel lateral cephalograms, which could, in time, reduce the need for radiation exposure to patients. The present study aimed to assess the accuracy of the Bellus3D ARC-7 (Bellus) camera by comparing the measurements of soft tissue facial landmarks obtained from Bellus scans to the measurements taken using Vernier callipers. Method Nineteen anatomical points were located on four subjects using a standard black ballpoint pen on a white, circular sticker. Distances were measured between these points using digital callipers, with the subject in a resting pose. This was repeated twice by two examiners for each subject. Two examiners subsequently performed measurements of each subject in a smiling pose. Following landmark identification, images were captured under standard conditions, using the Bellus camera. The same measurements were obtained digitally, repeated twice by two examiners for each subject in both resting and smiling poses. Results There was high precision in repeated measurements on the digital models, with less than 1.5 mm deviation between measurements. Both intra-examiner and inter-examiner reproducibility were greater following the digital measurements compared to manual measurements, with 100% of the digital measurements of landmarks falling within a set threshold deviation of ≤1.5 mm. When comparing the manual and digital measurements, the greatest deviations (>1.0 mm) occurred in regions around the cheeks and lower third of the face, while the measurements for the ears and midline structures (forehead and nose bridge) deviated the least (≤1.0 mm). This was demonstrated in models at rest and smiling. Conclusions The Bellus system produced an accurate and true image of the face from which reproducible measurements can be made within and between examiners. 3D facial images from the Bellus3D ARC-7 system were comparable to direct anthropometry, therefore the use of 3D facial scanning in orthodontics for diagnosis and treatment planning appears promising.

Publisher

Walter de Gruyter GmbH

Subject

Orthodontics

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3