In Vitro Microbiology Studies on a New Peritoneal Dialysis Connector

Author:

Di Bonaventura Giovanni12,Cerasoli Paolo3,Pompilio Arianna12,Arrizza Fabio4,Di Liberato Lorenzo5,Stingone Antonio6,Sirolli Vittorio5,Arduini Arduino4,Bonomini Mario5

Affiliation:

1. Center of Excellence on Ageing, 2 Clinical Microbiology Unit, “G. d'Annunzio” University, Chieti

2. Department of Biomedical Sciences, 2 Clinical Microbiology Unit, “G. d'Annunzio” University, Chieti

3. Nephrology and Dialysis Unit, Sulmona Hospital, Sulmona

4. R&D Department, Glomeria Therapeutics, Chieti

5. Department of Medicine, Institute of Nephrology, “G. d'Annunzio” University, Chieti

6. Nephrology and Dialysis Unit, Ortona Hospital, Ortona, Italy

Abstract

Objective We evaluated the ability of a recently developed peritoneal dialysis (PD) connector to prevent the risk of bacterial transfer to the fluid path after simulated touch and airborne contamination. Methods Staphylococcus epidermidis ATCC1228 and Pseudomonas aeruginosa ATCC27853 strains were used. For touch contamination, 2 μL of a standardized inoculum [1x108 colony-forming units (CFU) per milliliter] were deposited on top of the pin closing the fluid path of the patient connector. For airborne contamination, the patient connector was exposed for 15 seconds to a nebulized standardized inoculum. To simulate the patient peritoneum and effluent, the patient connector was pre-attached to a 2-L bag of sterile PD solution. After contamination, the patient connector was attached to the transfer set, the pin was captured, flow control was turned to simulate “patient drain” into the empty bag, and then “patient fill” using the bag pre-attached to the connector. Finally, a new pin was recaptured. The PD solution collected in the bag pre-attached to the connector was run through a 0.20-μm filter for colony counts. Results No infected connector transferred bacteria to the fluid path, regardless of the challenge procedure or the strain used. Conclusions Our results show that the new PD connector may fully obviate the risk of bacterial infection, even in the presence of heavy contamination. Further studies are in progress to test our PD connector in a clinical setting.

Publisher

SAGE Publications

Subject

Nephrology,General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3