Author:
Patnaik Jennifer L., ,Berdahl John P.,Hardten David R.,Wagner Brandie D.,Seibold Leonard K.,Kramer Brent A., , , , ,
Abstract
AIM: To assess the relationship between axial length (AL) and intraocular lens (IOL) rotation among eyes receiving a toric IOL and subsequently entered into an online toric back-calculator database.
METHODS: Retrospective analysis of data collected online via astigmatismfix.com, a freely available online toric back-calculator where surgeons enter pre- and post-operative information to help manage residual postoperative astigmatism. Included records were deemed valid with entry of AL and IOL orientation between January 2017 and March 2019. Rotation was determined by a difference of ≥5° between pre-operative intended IOL orientation and actual post-operative IOL orientation. Frequency and magnitude of rotation are presented with means and associated standard deviation (SD). Linear regression models of this association are presented.
RESULTS: Records of 6752 eyes were included in the analysis, of which 74.8% were determined to have a rotated IOL. The magnitude of rotation increased with each millimeter (mm) increase in AL with a mean rotation of 13.3° (SD: 12.8°) for eyes with AL 20-20.9 mm and a maximum mean rotation of 30.6° (SD: 30.3°) among eyes with AL 29-29.9 mm. General linear modeling demonstrated a significant association (P<0.0001) with a parameter estimate of 1.19 (standard error: 0.159) and R2 of 0.0083.
CONCLUSION: Analysis from an online database indicates that toric IOLs inserted into eyes with longer AL are more likely to rotate and to rotate more degrees from the target axis. The findings from this study are clinically relevant for surgeons implanting toric IOLs.
Publisher
Press of International Journal of Ophthalmology (IJO Press)
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献