Physical and clinical comparison between a screen-film system and a dual-side reading mammography-dedicated computed radiography system

Author:

Rivetti S.1,Canossi B.2,Battista R.2,Lanconelli N.3,Vetruccio E.2,Danielli C.4,Borasi G.5,Torricelli P.2

Affiliation:

1. Department of Health Physics, AUSL Modena, Modena, Italy

2. Department of Radiology, University of Modena and Reggio Emilia, Modena, Italy

3. Department of Physics, University of Bologna, Bologna, Italy

4. Department of Health Physics, Policlinico di Modena, Modena, Italy

5. Department of Health Physics, Arcispedale Santa Maria Nuova, Reggio Emilia, Italy

Abstract

Background: Digital mammography systems, thanks to a physical performance better than conventional screen-film units, have the potential of reducing the dose to patients, without decreasing the diagnostic accuracy. Purpose: To achieve a physical and clinical comparison between two systems: a screen-film plate and a dual-side computed radiography system (CRM; FUJIFILM FCR 5000 MA). Material and Methods: A unique feature of the FCR 5000 MA system is that it has a clear support medium, allowing light emitted during the scanning process to be detected on the "back" of the storage phosphor plate, considerably improving the system's efficiency. The system's physical performance was tested by means of a quantitative analysis, with calculation of the modulation transfer function, detective quantum efficiency, and contrast-detail analysis; subsequently, the results were compared with those achieved using a screen-film system (SFM; Eastmann Kodak MinR-MinR 2000). A receiver operating characteristic (ROC) analysis was then performed on 120 paired clinical images obtained in a craniocaudal projection with the conventional SFM system under standard exposure conditions and also with the CRM system working with a dose reduced by 35% (average breast thickness: 4.3 cm; mean glandular dose: 1.45 mGy). CRM clinical images were interpreted both in hard copy and in soft copy. Results: The ROC analysis revealed that the performances of the two systems (SFM and CRM with reduced dose) were similar ( P>0.05): the diagnostic accuracy of the two systems, when valued in terms of the area underneath the ROC curve, was found to be 0.74 for the SFM, 0.78 for the CRM (hard copy), and 0.79 for the CRM (soft copy). Conclusion: The outcome obtained from our experiments shows that the use of the dual-side CRM system is a very good alternative to the screen-film system.

Publisher

SAGE Publications

Subject

Radiology Nuclear Medicine and imaging,General Medicine,Radiological and Ultrasound Technology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3