Background/Aim The importance of an accurate determination of central corneal thickness (CCT) relies on its diagnostic and therapeutic implications in glaucoma, corneal ectasias,
corneal edema and endothelial function monitoring, and refractive surgery suitability, among others. We aimed to analyze the repeatability, reproducibility, correlation, and laterality
variations of CCT measurements performed with the Pentacam HR and the standard of care ultrasound A-scan (USP).
Methods A cross-sectional study including CCT measurements of healthy individuals was performed by three independent examiners with the Pentacam HR and USP. Intra-observer and
inter-observer variations were calculated with intra-class correlation coefficients (ICCs). Bland-Altman plots and 95% limits of agreement (95% LoA) were used to assess the agreement between
devices. Linear correlation was calculated with Pearsonʼs coefficient.
Results Thirty individuals (60 eyes), including 10 (33.3%) men and 20 (66.6%) women, with a mean age of 30.0 ± 9.1 years, were studied. No statistical differences were found in CCT
measurements between Pentacam HR (range 500 – 609 µm) and USP (range 498 – 628 µm). There was a high degree of correlation in repeatability and reproducibility of each independent device
(ICC > 0.90). Pearsonʼs correlation between 1 vs. 2, 2 vs. 3, and 3 vs. 1 Pentacam HR attempts were 0.914, 0.958, and 0.925, respectively (p < 0.001). Corresponding results for USP
were 0.957, 0.957, and 0.943 (p < 0.001). The Pentacam HR tended to overestimate CCT by a mean difference of 3.77 µm (95% LoA, − 24.9 – 18.4). Right eyes were also overestimated
(− 3.6 ± 14.1 µm) with the Pentacam HR device, whereas left eyes were underestimated (1.3 ± 11.1 µm).
Conclusions The Pentacam HR device provides reliable operator-independent estimates of CCT. Right eyes exhibited a tendency to overestimate with the Pentacam HR. We suspect this
difference is due to USP underestimation related to patientsʼ position while performing the study. In clinically relevant scenarios, performing a third measurement and cautiously measuring
right eyes can provide higher accuracy.