Performance of the microINR Point-of-Care System Used by Self-Testing Patients: A Multicenter Clinical Trial

Author:

Refaai Majed A.1,Jacobson Alan K.2,Rosenfeld Jack C.3,Orr Robert R.4

Affiliation:

1. Department of Pathology and Laboratory Medicine, University of Rochester Medical Center, Rochester, New York, United States

2. Department of Internal Medicine, VA Loma Linda Healthcare System, Loma Linda, California, United States

3. Green and Seidner Family Practice Associates, Lansdale, Pennsylvania, United States

4. Phoenix Medical Research, Peoria, Arizona, United States

Abstract

Abstract Introduction Anticoagulation monitoring is a major practical and clinical challenge. We assessed the performance of the microINR system in patient self-testing (PST). Methods This study was performed at four US medical centers. After the training visit of warfarin anticoagulated patients (n = 117) on microINR system, PST was performed at home and in two visits to the medical centers. At the medical centers, both PST and healthcare professionals (HCPs) performed duplicate tests with the microINR System. A venous blood sample for the laboratory testing was also extracted. Accuracy and precision were assessed. Results The comparison between microINR PST results and microINR HCP results revealed an equivalence with a slope of 1.00 (95% confidence interval [CI]: 1.00–1.00), and an intercept of 0.00 (95% CI: 0.00–0.00). When compared with the laboratory analyzer, microINR PST results also showed good correlation with a slope of 0.94 (95% CI: 0.86–1.04) and an intercept of 0.14 (95% CI: -0.09–0.34). Predicted bias values at international normalized ratio (INR) 2.0, 3.5, and 4.5 were 0% against HCP and ≤2.5% against the laboratory. Analytical agreement with both HCP and laboratory was 100% according to ISO17593 and 99.1 and 100% according to CLSI POCT14 with HCP and laboratory, respectively. Clinical agreement with HCP regarding 2.0–4.0 as INR therapeutic range was 98% (within range). The precision (coefficient of variation) of microINR system used by PST was comparable to HCP. Conclusion The microINR results when used by self-testing patients show satisfactory concordance to both HCP results and laboratory analyzer. The microINR system is adequate for self-testing use.

Publisher

Georg Thieme Verlag KG

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3