Affiliation:
1. Department of Biochemistry, Mahatma Gandhi Medical College and Research Institute, Sri Balaji Vidyapeeth (Deemed to be University), Pondicherry, India
Abstract
Abstract
Introduction With increasing automation in clinical laboratories, the requirements for quality control (QC) material have greatly increased in order to monitor performance. The constant use of commercial control material is not economically feasible for many countries because of nonavailability or the high-cost of those materials. Therefore, preparation and use of in-house QC serum will be a very cost-effective measure with respect to laboratory needs.
Materials and Methods In-house internal quality control from leftover serum samples of master health checkup subjects, which have been screened negative for HIV, HCV and HBsAg antibodies was pooled in a glass jar with ethanediol as preservative and kept in deep freezer at − 20°C. From the pooled serum, 100 microliter thirty aliquots were prepared. Every day along with commercial internal QC (IQC), one aliquot of pooled serum was analyzed for 30 days for the following parameters: plasma glucose, blood urea, serum creatinine, total cholesterol, triglycerides (TGL), high-density lipoprotein, calcium, total protein, albumin, total bilirubin, AST, ALT, ALP, amylase. After getting 30 values for each parameter, mean, standard deviation (SD) and CV% were calculated for both IQC commercial sample and pooled serum sample.
Results The mean, SD, and CV% of glucose, cholesterol, TGL, calcium, alanine aminotransaminase (ALT), aspartate aminotransferase (AST), amylase, and alkaline phosphatase (ALP) were statistically significant between pooled serum and commercial QC.
Conclusion In-house QC prepared from pooled serum is better than commercial internal QC. The biochemical parameters were stable in pooled serum due to less matrix effect; also, variation was less in pooled serum IQC.
Reference12 articles.
1. Quality control in clinical biochemistry;A S Kanagasabapathy;Indian J Clin Biochem,1996
2. Quality control for the clinical chemistry laboratory;S Ohman;Qual Assur,1997
3. Ethanediol stabilised quality control serum for measurement of zinc;S Swaminathan;Clin Chim Acta,1993
4. External quality assurance;J Middle;Ann Clin Biochem,1998
5. Economy and quality assessment of home made clinical chemistry reagents;M N Zafar;J Pak Med Assoc,1992
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献