Abstract Objective The quantity of A-trains, a high-frequency pattern of free-running facial nerve electromyography, is correlated with the risk for postoperative high-grade facial nerve paresis. This correlation has been confirmed by automated analysis with dedicated algorithms and by visual offline analysis but not by audiovisual real-time analysis. Methods An investigator was presented with 29 complete data sets measured during actual surgeries in real time and without breaks in a random order. Data were presented either strictly via loudspeaker (audio) or simultaneously by loudspeaker and computer screen (audiovisual). Visible and/or audible A-train activity was then quantified by the investigator with the computerized equivalent of a stopwatch. The same data were also analyzed with quantification of A-trains by automated algorithms. Results Automated (auto) traintime (TT), known to be a small, yet highly representative fraction of overall A-train activity, ranged from 0.01 to 10.86 s (median: 0.58 s). In contrast, audio-TT ranged from 0 to 1,357.44 s (median: 29.69 s), and audiovisual-TT ranged from 0 to 786.57 s (median: 46.19 s). All three modalities were correlated to each other in a highly significant way. Likewise, all three modalities correlated significantly with the extent of postoperative facial paresis. As a rule of thumb, patients with visible/audible A-train activity < 1 minute presented with a more favorable clinical outcome than patients with > 1 minute of A-train activity. Conclusion Detection and even quantification of A-trains is technically possible not only with intraoperative automated real-time calculation or postoperative visual offline analysis, but also with very basic monitoring equipment and real-time good quality audiovisual analysis. However, the investigator found audiovisual real-time-analysis to be very demanding; thus tools for automated quantification can be very helpful in this respect.

Funder

Deutsche Forschungsgemeinschaft DFG

Publisher

Georg Thieme Verlag KG

Subject

Neurology (clinical),Surgery

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3