Clinical Short-Term Outcome and Hemodynamic Comparison of Six Contemporary Bovine Aortic Valve Prostheses

Author:

Hartrumpf Martin1,Kuehnel Ralf-Uwe1,Schroeter Filip1,Haase Robert1,Laux Magdalena L.1,Ostovar Roya1,Albes Johannes M.1

Affiliation:

1. Department of Cardiovascular Surgery, Heart Center Brandenburg, Brandenburg Medical School, University Hospital Bernau, Bernau bei Berlin, Germany

Abstract

Abstract Background Conventional stented valves (CV) remain gold standard for aortic valve disease. Bovine prostheses have been improved and rapid deployment valves (RDV) have arrived in the recent decade. We compare clinical and hemodynamic short-term outcome of six bovine valves. Methods We retrospectively evaluated 829 consecutive patients (all-comers) receiving bovine aortic valve replacement (AVR). Four CV from different manufacturers (Mitroflow, Crown, Perimount, Trifecta) and two RDV (Perceval, Intuity) were compared in terms of pre-, intra-, and postprocedural data. A risk model for mortality was created. Results All valves reduced gradients. From 23 mm, all CV showed acceptable gradients. Twenty-one millimeter Mitroflow/Perceval and 19 mm Crown showed above-average gradients. As baseline data differed, we performed propensity matching between aggregated isolated CV and RDV groups. Cardiopulmonary bypass (CPB), clamp, and surgery times were shorter with RDV (87.4 ± 34.0 min vs 111.0 ± 34.2, 54.3 ± 21.1 vs 74.9 ± 20.4, 155.2 ± 42.9 vs 178.0 ± 46.8, p < 0.001). New pacemaker rate (10.1 vs 1.3%, p = 0.016) and the tendency toward neurologic events (8.9 vs 2.5%, p = 0.086) were higher using RDV, induced mainly by the Perceval. Early mortality was equal (2.5 vs 1.3%, p = 0.560). Revision for bleeding, dialysis, blood products, length-of-stay, gradients, and regurgitation was also equal. Risk analysis showed that low valve size, low ejection fraction, endocarditis, administration of red cells, and prolonged CPB time were predictors of elevated mortality. Conclusion Isolated bovine AVR has low mortality. Valves ≥ 23 mm show comparable gradients while the valve model matters < 23 mm. RDV should be used with care. Procedure-related times are shorter than those of CV but pacemaker implantation and neurologic events are more frequent (Perceval). Early mortality is low and valve performance comparable to CV.

Publisher

Georg Thieme Verlag KG

Subject

Cardiology and Cardiovascular Medicine,Pulmonary and Respiratory Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3