Evaluation of a New Rapid Quantitative D-dimer Assay in Patients with Clinically Suspected Deep Vein Thrombosis

Author:

D’Angelo Armando1,D’Alessandro Gabriella1,Tomassini Loredana1,Pittet Jean Louis2,Dupuy G2,Crippa Luciano1

Affiliation:

1. The Servizio di Coagulazione, Istituto Scientifico H. S. Raffaele, Milano, Italy

2. Laboratoire d’Hemostase, bioMérieux, Marcy-l’Etoile, France

Abstract

SummaryThe sensitivity and specificity for deep vein thrombosis (DVT) of a new rapid, quantitative and precise (total imprecision < 10%) D-dimer assay suitable for individual measurements (VIDAS D-DIMER, bio-Merieux, France) were evaluated in a consecutive series of 103 in- and out-patients submitted to serial compression ultrasonography (C-US) for the clinical suspicion of DVT (n = 66) or of DVT recurrence (n = 37) and symptoms lasting from 1 to 15 days. DVT was found in 22 patients at baseline testing and no patient with an initially negative C-US developed vein incompressibility at follow up. The time elapsed from the onset of symptoms was negatively associated with D-dimer levels both in patients with and in those without DVT. In the entire series of patients, the sensitivity of a positive D-dimer test (≥1.0 Μg/ml) for the presence of DVT was 96% (21/22 patients, 95% confidence interval 75-100%) with a specificity of 75% (64-84%), a negative predictive value of 98% (90-100%), a positive predictive value of 51% (35-67%), and an overall accuracy of 80% (70-87%). A normal D-dimer value (0.22 Μg/ml) was observed in one patient with DVT and symptoms lasting from 15 days. The approach of withholding C-US testing in patients with symptoms lasting from less than 11 days and D-dimer levels below the cut-off value was compared to serial C-US testing alone in a cost-effectiveness analysis subdividing the 66 patients with a first episode according to their clinical pretest probability of DVT. Thrombosis was detected in 6.7% of the patients in the low probability group (n = 15), 16.7% of the patients in the moderate probability group (n = 24), 51.9% of the patients in the high probability group (n = 27) and 8.1% of patients with suspected DVT recurrence. Calculated cost-savings for each DVT diagnosed ranged from 5% in the high pretest probability group to 55% in the low pretest probability group and to 77% in patients with suspected DVT recurrence.The safety of avoiding C-US testing in symptomatic patients with a negative D-dimer test should be evaluated in clinical management studies.

Publisher

Georg Thieme Verlag KG

Subject

Hematology

Cited by 73 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3