Affiliation:
1. The Haemostasis Department, Clinical Laboratory, St. Elisabeth’s Hospital, Arnhem The Netherlands
Abstract
SummarySix commercially available reagents for the determination of the activated partial thromboplastin time have been evaluated and compared with respect to their sensitivity to the coagulation factors VIII, IX and XI and to their response to heparin. Some variation was observed among the reagents regarding their sensitivity to factor XI and even greater differences were obtained with factors VIII and IX. It was also clear that none of the reagents was sensitive to the same extent to the factors tested. The sensitivity to heparin shows considerable variation, in terms of time as well as mode of response to increasing heparin levels. In four reagents this response is linear, it is logarithmic in one and the remaining one is yet again different.It seems unlikely that any standardization of the APTT determination is at present possible with the reagents studied.
Cited by
26 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献