Background Optimal surgical treatment of functional ischemic mitral regurgitation (FIMR) is still controversy. Due to the underlying pathophysiology, stand-alone ring annuloplasty is assumed with a high recurrence rate of mitral regurgitation, thus additional subvalvular repair techniques might improve the results. This in vitro study introduces a new device for subvalvular mitral valve repair.
Methods We developed a new device for subvalvular mitral valve repair, consisting of two frames for papillary muscle (PM) attachment, which are connected with two holding bars serving for fixation of the device on an annuloplasty ring. In the first experimental run, porcine mitral valves including the chordae tendineae and PMs were fixated on a holding device, consisting of a holding ring simulating mitral annulus dilation and height-adjustable frames for PM attachment simulating leaflet tethering. In vitro regurgitant volume was determined in a pulse duplicator. Afterward, the frames for PM attachment were replaced by our newly developed device and the measurements were repeated.
Results In the model simulating FIMR, the regurgitant volume was 44.3 ± 12.38 mL/stroke. After subvalvular reconstruction with our new device, the regurgitant volume was significantly reduced to 33.1 ± 11.68 mL/stroke (p = 0.009).
Conclusion In this specific in vitro model, our new device for subvalvular mitral valve reconstruction led to a significant reduction of the regurgitant volume, thus representing a promising technique to potentially improve the results of mitral reconstruction in ischemic functional mitral valve regurgitation. Additional studies are required to further investigate and improve our device.
Subject
Cardiology and Cardiovascular Medicine,Pulmonary and Respiratory Medicine,Surgery
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献