Monitoring Heparin Therapy Using Activated Partial Thromboplastin Time – Results of a Multicenter Trial Establishing the Therapeutic Range for SILIMAT, a Reagent with High Sensitivity to Heparin

Author:

Boutière Brigitte,Horellou Marie-Hélène,Trzeciak Marie Christine,Samama Meyer,Toulon Pierre

Abstract

SummaryAPTT is widely used for laboratory monitoring of treatment with unfractionated heparin (UFH). However, since its sensitivity to heparin varies significantly from one reagent to another, the therapeutic range had to be defined for each brand of APTT reagent. As an example, SILIMAT (bio-Mérieux) is a new APTT reagent containing rabbit brain phospholipids and micronized silica as an activator. Since its high sensitivity to heparin has been previously reported, a multicenter trial was carried out in an attempt to define the therapeutic range of APTT performed using this new reagent. For that purpose, 170 blood samples drawn for routine coagulation testing from 170 different patients treated with UFH were analyzed. A single batch of two different APTT reagents were used on KC10 coagulometers: SILIMAT and Automated APTT (Organon-Teknika) whereas the anti-Xa activity was evaluated by a chromogenic substrate-based assay. The same methodology was used in all the centers. In order to obtain a plasma anti-Xa activity within the therapeutic range i.e. between 0.30 and 0.70 IU/ml, the APTT ratios were found between 1.90 and 5.40 for SILIMAT, which corresponded to clotting times of the patients plasma between 63 and 178 s. The APTT ratios were significantly lower when evaluated using Automated APTT (between 1.70 and 4.10), with clotting times between 53 and 127 s. In addition, a good correlation was found between the Anti-Xa activity and APTT for both reagents (r >0.65). However, it is not possible to make recommendations regarding the therapeutic ranges without restrictions. Although about 70% of the patients with an anti-Xa activity between 0.30 and 0.70 IU/ml had an APTT in the above defined ranges, the degree of concordance between the two assays is not absolute. Actually more than 30% of the patients had discordant anti-Xa activity and APTT and more than a quarter of the patients included in the above defined therapeutic range for APTT had an anti-Xa activity outside the 0.30-0.70 IU/ml range, whatever the reagent used. In conclusion, to define the therapeutic ranges of APTT using the recommended method is practicable but some critical points could be raised, suggesting that a better method is awaited in order to improve the standardization.

Publisher

Georg Thieme Verlag KG

Subject

Hematology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3