Assessment of the Influence of Citrate Concentration on the International Normalized Ratio (INR) Determined with Twelve Reagent-instrument Combinations

Author:

Chantarangkul Veena,Clerici Marigrazia,Negri Barbara,Mannucci Pier Mannuccio,Tripodi Armando

Abstract

SummaryCitrate concentration is one of the variables that can affect coagulation tests. However, few studies have so far been performed to assess the magnitude of this effect on coagulation tests in general and PT in particular. The aim of this study was to assess the extent of influence of citrate concentration on the PT test with results expressed as INR. Twelve reagent-instrument combinations (systems) were calibrated vs. the Reference Preparation BCT/441 using plasmas collected in either 105 mM or 129 mM citrate from normals and anticoagulated patients (OAT). PTs for plasmas collected in 129 mM citrate were longer than those collected in 105 mM both for normals and patients on OAT, but the ratios (patient-to-normal clotting times) for the two citrate concentrations were significantly different in many instances, implying that the International Sensitivity Index (ISI) is also different. ISIs for calibrations with plasmas collected in 105 mM were greater (up to 10%) than those with plasmas collected in 129 mM citrate. When PT ratios were transformed into INR using crossover ISIs (i.e., plasmas collected in 105 mM and ISI determined with plasmas collected in 129 mM citrate, or vice versa) we found that an INR of 4.5 could be up to 20% apart from the value that would have been obtained if the appropriate ISI was used. Moreover, if the ISI determined with the manual technique was used to convert PTs obtained with a particular instrument into INR, the effect of citrate concentration was even greater (INR difference up to 64%). Should these observations be valid for other systems, they might provide additional explanations for the frequent reports which document discrepancies in the INR determined with different systems to which incorrect ISI might have been applied. World-wide consensus on a single citrate concentration to collect patients’ as well as lyophilized plasmas to be used in External Quality Assessment Schemes and for local system calibration is therefore urgently needed.

Publisher

Georg Thieme Verlag KG

Subject

Hematology

Cited by 29 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3