Prospective Comparison of a Metal-Free Ceramic Total Knee Arthroplasty with an Identical Metal System

Author:

Trieb Klemens1,Ullmann David1,Metzinger Katharina1,Kramar Reinhard1,Mauerer Andreas23,Hofstätter Stefan1,Fiala Rainer1

Affiliation:

1. Klinikum Wels-Grieskirchen, Abteilung für Orthopädie und Orthopädische Chirurgie, Wels, Austria

2. Klinik für Unfallchirurgie und Orthopädie, St. Theresien Krankenhaus, Nürnberg, Germany

3. Biomechanisches Labor – UO Lab, Universitätsklinikum, Erlangen, Germany

Abstract

Abstract Aims The aim of this open-label, prospective, short-term study was to carry out an initial comparison of a completely metal-free ceramic with a geometrically identical metallic arthroplasty over a 1-year follow-up period. Methods This study investigates a completely metal-free system using a composite matrix material containing aluminum oxide (Al2O3, BPK-S Integration, Peter Brehm GmbH, Weisendorf, Germany) or zirconium oxide (ZrO2, BPK-S Integration Ceramic, Biolox Delta-CeramTec GmbH, Plochingen, Germany). Eighty patients (40 in each group) received either a completely metal-free ceramic system (matrix of aluminum and zirconium oxide) or an anatomically identical metallic knee system made of a cobalt-chromium alloy. Clinical assessment was performed preoperatively, and during follow-up at 3 and 12 months, using the Knee Society Score, Oxford Knee Score, and EQ-5D-VAS. For radiological evaluation, standard preoperative and postoperative standardized radiographs were taken at the given follow-up visits. Results The postoperative clinical scores improved significantly at the 3- and 12-month follow-ups, but did not differ statistically between the two groups. The radiologically evaluated mean postoperative mechanical and anatomical axes showed proper alignment within both groups at all times. No revision surgery had to be performed, and no complications or loosening were recorded whatsoever. Conclusions To our knowledge, this is the first study comparing a total ceramic metal-free knee system with a geometrically identical metallic TKA (total knee arthroplasty). Longer follow-ups will be required to demonstrate the overall efficiency of this TKR and perhaps to expand its medical indication.

Publisher

Georg Thieme Verlag KG

Subject

Orthopedics and Sports Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3