Affiliation:
1. Department of Medical and Occupational Radiation Protection, Federal Office for Radiation Protection, Neuherberg, Germany
2. Institute for Radiology and Nuclear Medicine, Paracelsus Medical University, General Hospital Nuremberg, Germany
3. Diagnostic and Interventional Radiology, Ruhr-Universität Bochum, Germany
Abstract
Purpose Recent developments in medical technology have broadened the spectrum of X-ray procedures and changed exposure practice in X-ray facilities. For this reason, diagnostic reference levels (DRLs) for diagnostic and interventional X-ray procedures were updated in 2016 and 2018, respectively. It is the aim of this paper to present the procedure for the update of the DRLs and to give advice on their practical application.
Materials and Methods For the determination of DRLs, data from different independent sources that collect dose-relevant data from different facilities in Germany were considered. Seven different weight intervals were specified for classifying pediatric X-ray procedures. For each X-ray procedure considered, the 25th, 50th, and 75th percentile of the respective national distribution of the dose-relevant parameters were determined. Additionally, effective doses that correspond to the DRLs were estimated.
Results In procedures with already existing DRLs before 2016, the values were lowered by circa 20 % on average. Numerous DRLs were established for the first time (9 for interventional procedures, 10 for CT examinations).
Conclusion For dose optimizations even below the new national DRLs, the BfS recommends establishing local reference levels, using dose management software (particularly in CT and interventional radiology), adapting dose-relevant parameters of X-ray protocols to the individual patient size, and establishing internal radiation protection teams responsible for optimizing X-ray procedures in clinical practice. When applying good medical practice and using modern equipment, the median dose values of the nationwide dose distributions can not only be easily achieved but can even be undercut.
Key Points:
Citation Format
Subject
Radiology, Nuclear Medicine and imaging
Cited by
102 articles.