A Comparison of Optical Biometers Used in Children for Myopia Control

Author:

Mattern Ann-Isabel1,Neller Kai12ORCID,Devenijn Machteld1,Schwahn Hartmut1,Langenbucher Achim2,Seitz Berthold3ORCID,Kaymak Hakan12ORCID

Affiliation:

1. Internationale Innovative Ophthalmochirurgie GbR c/o Breyer Kaymak and Klabe Augenchirurgie, Düsseldorf, Germany

2. Institute of Experimental Ophthalmology, Saarland University, Homburg, Germany

3. Dept. of Ophthalmology, Saarland University Hospital and Faculty of Medicine, Homburg, Germany

Abstract

Abstract Purpose To assess the reproducibility (i.e., inter-device reliability) of the biometers Topcon MYAH, Oculus Myopia Master, and Haag-Streit Lenstar LS900 with the Carl Zeiss IOLMaster 700 and the intra-subject repeatability in myopic children in order to reliably detect axial growth for myopia management. Methods Twenty-two myopic children (11.1 ± 2.4 yr) with a spherical equivalent of − 3.53 ± 2.35 D were examined with each of the biometers to assess axial length (AL) and corneal parameters (steepK, flatK, meanK, vectors J0, J45), and16 of these children agreed to a second round of measurements. Reproducibility of the first measurements between the IOLMaster and every other biometer was assessed employing a Bland-Altman approach and paired Studentʼs t-test. Repeatability was assessed as intra-subject standard deviation and was used to estimate the minimum time interval required between two AL measurements to reliably detect axial growth of an eye of at least 0.1 mm/year. Results Repeatability for AL measurements was as follows: IOLMaster: 0.05 mm, Myopia Master: 0.06 mm, Myah: 0.06 mm, Lenstar: 0.04 mm; the respective minimal time interval for axial growth assessment in myopia management was estimated as 5.6, 6.6, 6.7, and 5.0 months, respectively. Best reproducibility of the AL measurement was found between IOLMaster and Lenstar [95% Limits of Agreement (LoA) for reproducibility − 0.06 to 0.02]. As regards to the measured means, Lenstar gave measures of AL that were longer than with the IOLMaster by 0.02 mm (p < 0.001). Myopia Master measures of meanK were significantly lower (by 0.21 D with p < 0.001) than the values from the IOLMaster. As for J0, all biometers deviated significantly from IOLMaster measurements (p < 0.05). Conclusion Generally good agreement was observed between all the biometers. When assessing myopia progression in children, a time frame of at least 6 months between the AL measurements is advisable in order to reliably determine any deviation from a normal growth pattern.

Publisher

Georg Thieme Verlag KG

Subject

Ophthalmology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3