Safety and Efficacy of Overdilation of 10 mm Viatorr Transjugular Intrahepatic Portosystemic Shunt Stents Using 12 mm Balloons

Author:

Tahir Muhammad Mohid1ORCID,Lewis Trevor K.1,Ali Aamir1,Hsu Michael1,Weinstein Jeffrey L.1,Ahmed Muneeb1,Sarwar Ammar1

Affiliation:

1. Department of Radiology, Beth Israel Deaconess Medical Center, Boston, Massachusetts, United States

Abstract

Abstract Objective The aim of this study was to evaluate overdilation of 10-mm standard and 8–10 mm controlled-expansion Viatorr stents to 12 mm during transjugular intrahepatic portosystemic shunt (TIPS) placement when insufficient reduction (<50%) in portosystemic gradient (PSG) is achieved with standard 10 mm dilation. Materials and Methods It is a single-institution, institutional review board-approved, retrospective review of TIPS (2013–2022) to identify patients in the overdilation group (12 mm dilation of a 10-mm stent) and a control group (10 mm dilation of a 10-mm stent) matched for age, indication, stent type, Model for End-Stage Liver Disease (MELD) score, pre-TIPS PSG, and variceal embolization. Stent diameter, technical success, clinical outcomes, and adverse events were assessed for both groups. Results TIPS was created for the overdilation group (n = 35, 57 ± 11 years, 69% male; MELD: 14 ± 5) and control group (n = 35, 57 ± 11 years, 83% male; MELD: 14 ± 5). Overdilation to 12 mm adequately reduced PSG by more than 50% (55 vs. 65% in the control group, p = 0.11). The stent diameter was larger in the overdilation group on cross-sectional imaging (9.8 ± 0.2 vs. 9.5 ± 0.4 mm, p < 0.001), with an estimated 57% higher volume flow rate (p = 0.002). Patients were followed for a median of 11.3 months (range: 0.03–75) and 15.6 months (range: 0.03–106) in the overdilation and control groups, respectively. There was an equivalent rate of ascites resolution (56 vs. 63%, p = 0.68) and rebleeding (13 vs. 17%, p = 0.82) in the overdilation and control groups, with a similar risk of new-onset hepatic encephalopathy (41 vs. 33%, p = 0.51) and TIPS occlusion (11 vs. 9%, p = 0.69). Overdilation did not result in any instance of stent fracture. Conclusion Overdilation of 10-mm Viatorr stents with 12 mm balloons may provide benefit by potentially reducing PSG further for patients initially having inadequate PSG reduction with short-term safety.

Publisher

Georg Thieme Verlag KG

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3