No relation between adenosine triphosphate after manual cleaning and presence of microorganisms on endoscopes after automated high-level disinfection

Author:

Vos Margreet C.1,Bruno Marco J.2,Kwakman Judith A.21,Rauwers Arjan W.21,Buijs Jolanda G.3,de Groot Woutrinus1

Affiliation:

1. Department of Medical Microbiology and Infectious Diseases, Erasmus MC University Medical Center, Rotterdam, the Netherlands

2. Department of Gastroenterology and Hepatology, Erasmus MC University Medical Center, Rotterdam, the Netherlands

3. Qualtity Assurance and Regulatory Affairs Office Medical Devices, Erasmus MC University Medical Center, Rotterdam, the Netherlands

Abstract

Abstract Background and study aims Adenosine triphosphate (ATP) tests are increasingly used to detect biological material; however, their reliability to detect bacterial contamination in endoscopes is not proven. We investigated the predictive value of ATP tests after manual cleaning for presence or absence of microorganisms as shown by culture after automated high-level disinfection (HLD) in duodenoscopes and linear echoendoscopes (DLEs). Patients and methods After manual cleaning, ATP tests were performed on swab samples taken from the detachable cap and forceps elevator, and on flush samples of the DLE working channels. These results were compared to the growth of any microorganisms in cultures acquired after automated HLD. ATP tests with > 200 relative light units (RLU) were considered positive. Receiver operator characteristic (ROC) curves were used to compare the RLU levels with microbial presence in cultures. Results In total, 903 procedures were performed involving 26 distinct DLEs. Depending on sample site, 20.8 % (cap) to 63.8 % (channel brush) of the ATP negative samples were accompanied by positive post-HLD cultures. 54.4 % of the cap samples with a positive culture (growth of any kind of microorganism) and 91.8 % of the channel samples with a positive culture had a negative ATP test after manual cleaning. ROC curves per sample site, DLE type and microorganism type all had area under the curves below 0.6. Conclusions In our study, ATP tests performed after manual cleaning could not predict presence or absence of microorganisms after automated HLD as shown by culture. More than half of the positive cultures were preceded by a negative ATP test.

Funder

3M Health Care

Publisher

Georg Thieme Verlag KG

Subject

Gastroenterology,Medicine (miscellaneous)

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3