Investigation of the efficacy of an innovative endoscope drying and storage method in a simulated ERCP setting

Author:

Vos Margreet C.1,Bruno Marco J.2,Kwakman Judith A.21

Affiliation:

1. Department of Medical Microbiology and Infectious Diseases, Erasmus Medical Center, Rotterdam, the Netherlands

2. Department of Gastroenterology and Hepatology, Erasmus Medical Center, Rotterdam, the Netherlands

Abstract

Abstract Background and study aims Drying after cleaning and disinfection is essential in the reprocessing of endoscopes since microorganisms can grow and form biofilms on wet surfaces. In this experimental non-clinical study, we investigated the efficacy of a novel, fast-drying method when reprocessing duodenoscopes. Methods During a series of 40 tests, three duodenoscopes were exposed to an artificial test soil containing supraphysiological loads of four types of gut microorganisms in a non-clinical ERCP simulation, followed by reprocessing and drying with the PlasmaTYPHOON. Cultures of the distal tip and working channel were acquired immediately after automated decontamination and after drying with the PlasmaTYPHOON. Cobalt chloride paper tests and borescope inspections were used to evaluate drying efficacy. Results Contamination of the working channels dropped from 86.4 % post-decontamination to 33.6 % post-drying, with 94 % of the positive post-drying samples belonging to one duodenoscope. This duodenoscope showed persistent contamination with P. aeruginosa in the working channel. The other two duodenoscopes only showed low levels of P. aeruginosa in post-decontamination channel samples, but not after drying. Cobalt chloride paper tests and borescope inspections revealed good drying efficacy. Conclusions Positive cultures for gut microorganisms were often found in wet endoscopes post-decontamination. The PlasmaTYPHOON is an effective fast-drying method capable of abolishing nearly all remaining microorganisms after decontamination provided no biofilm has developed, even when using a supraphysiological concentration of bacterial load. The clinical use of the PlasmaTYPHOON has the potential to reduce endoscope contamination, the use of wet contaminated endoscopes and therefore the risk of patient infection.

Publisher

Georg Thieme Verlag KG

Subject

Obstetrics and Gynecology

Reference20 articles.

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3