Intramuscular Microvascular Flow Sensing for Flap Monitoring in a Porcine Model of Arterial and Venous Occlusion

Author:

Lu Di1,Moritz William2,Arafa Hany M.1,Yang Quansan1,Jacobson Lauren2,Ostojich Diana1,Bai Wubin3,Guo Hexia1,Wu Changsheng1,Li Shuo1,Li Shupeng4,Huang Yonggang4,Xu Yameng5,Yan Ying5,Westman Amanda M.2,MacEwan Matthew R.5,Rogers John A.1,Pet Mitchell A.2ORCID

Affiliation:

1. Center of Bio-Integrated Electronics, Querrey Simpson Institute for Bioelectronics, Northwestern University, Evanston, Illinois

2. Division of Plastic and Reconstructive Surgery, Department of Surgery, School of Medicine, Washington University, St. Louis, Missouri

3. Department of Applied Physical Sciences, The University of North Carolina at Chapel Hill, Chapel Hill, North Carolina

4. Department of Mechanical Engineering, Northwestern University, Evanston, Illinois

5. Department of Neurosurgery, School of Medicine, Washington University, St. Louis, Missouri

Abstract

Abstract Background Commercially available near infrared spectroscopy devices for continuous free flap tissue oxygenation (StO2) monitoring can only be used on flaps with a cutaneous component. Additionally, differences in skin quality and pigmentation may alter StO2 measurements. Here, we present a novel implantable heat convection probe that measures microvascular blood flow for peripheral monitoring of free flaps, and is not subject to the same issues that limit the clinical utility of near-infrared spectroscopy. Methods The intratissue microvascular flow-sensing device includes a resistive heater, 4 thermistors, a small battery, and a Bluetooth chip, which allows connection to a smart device. Convection of applied heat is measured and mathematically transformed into a measurement of blood flow velocity. This was tested alongside Vioptix T.Ox in a porcine rectus abdominis myocutaneous flap model of arterial and venous occlusion. After flap elevation, the thermal device was deployed intramuscularly, and the cutaneous T.Ox device was applied. Acland clamps were alternately applied to the flap artery and veins to achieve 15 minutes periods of flap ischemia and congestion with a 15 minutes intervening recovery period. In total, five devices were tested in three flaps in three separate pigs over 16 vaso-occlusive events. Results Flow measurements were responsive to both ischemia and congestion, and returned to baseline during recovery periods. Flow measurements corresponded closely with measured StO2. Cross-correlation at zero lag showed agreement between these two sensing modalities. Two novel devices tested simultaneously on the same flap showed only minor variations in flow measurements. Conclusion This novel probe is capable of detecting changes in tissue microcirculatory blood flow. This device performed well in a swine model of flap ischemia and congestion, and shows promise as a potentially useful clinical tool. Future studies will investigate performance in fasciocutaneous flaps and characterize longevity of the device over a period of several days.

Publisher

Georg Thieme Verlag KG

Subject

Surgery

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3