Intraoperative Ergonomic Assessment of Exoscopes versus Conventional DIEP Flap

Author:

Wang Tianke1,Norasi Hamid12,Nguyen Minh-Doan3,Harless Christin3,Law Katherine E.12,Smith Tianqi G.1,Tetteh Emmanuel12,Hallbeck M. Susan123

Affiliation:

1. Robert D. and Patricia E. Kern Center for the Science of Healthcare Delivery, Mayo Clinic, Rochester, Minnesota

2. Division of Healthcare Delivery Research, Mayo Clinic, Rochester, Minnesota

3. Department of Surgery, Mayo Clinic, Rochester, Minnesota

Abstract

Abstract Background This study compared the ergonomics of surgeons during deep inferior epigastric perforator (DIEP) flap surgery using either baseline equipment (loupes, headlights, and an operating microscope) or an exoscope. Plastic surgeons may be at high risk of musculoskeletal problems. Recent studies indicate that adopting an exoscope may significantly improve surgeon postures and ergonomics. Methods Postural exposures, using inertial measurement units at the neck, torso, and shoulders, were calculated in addition to the surgeons' subjective physical and cognitive workload. An ergonomic risk score on a scale of 1 (lowest) to 4 (highest) was calculated for each of the postures observed. Data from 23 bilateral DIEP flap surgeries (10 baseline and 13 exoscope) were collected. Results The neck and torso risk scores decreased significantly during abdominal flap harvest and chest dissection, while right shoulder risk scores increased during the abdominal flap harvest for exoscope DIEP flap procedures compared with. Exoscope anastomoses demonstrated higher neck, right shoulder, and left shoulder risk scores. The results from the survey for the “surgeon at abdomen” showed that the usage of exoscopes was associated with decreased performance and increased mental demand, temporal demand, and effort. However, the results from the “surgeon at chest” showed that the usage of exoscopes was associated with lower physical demand and fatigue, potentially due to differences in surgeon preference. Conclusion Our study revealed some objective evidence for the ergonomic benefits of exoscope; however, this is dependent on the tasks the surgeon is performing. Additionally, personal preferences may be an important factor to be considered in the ergonomic evaluation of the exoscope.

Publisher

Georg Thieme Verlag KG

Subject

Surgery

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3