Affiliation:
1. Diagnostic Imaging Department, Davies Veterinary Specialists, Hitchin, United Kingdom
2. Willows Referral Centre, Solihull, United Kingdom
Abstract
Abstract
Objectives (1) To compare the ability of standard computed tomography (CT) scale (SCTS) and extended CT scale (ECTS) images, produced using conventional CT technology, to provide detailed assessment of metal screws in vitro. (2) To assess how screw size, type, and orientation relative to the Z-axis of the gantry affect implant assessment. (3) To test the ability of SCTS, ECTS, and radiography to diagnose screw failure when there is negligible screw fragment displacement.
Study Design Part 1: 12 screws of different size, type, and composition were scanned in three orientations (parallel or 0°; oblique or 45°; and perpendicular or 90°) relative to the Z-axis of the gantry. SCTS and ECTS reconstructions were made for each screw, in each plane, to assess implant shape, structure, and diameter. Part 2: fatigue-failure was induced in four screws commonly used to stabilize canine humeral intracondylar fissures. Screws were then reassembled achieving grossly perfect apposition and alignment of the fragments. Ability to detect implant failure was tested using SCTS, ECTS, and radiography.
Results ECTS provided better screw assessment compared with SCTS resulting in clear visualization of the structure in 8/12 versus 0/12 screws and shape in 12/12 versus 11/12 screws; however, results were affected by screw size, type, and orientation. ECTS identified all in vitro screw fractures with negligible screw fragment displacement; however, success was affected by screw orientation: 4/4 fractures identified with a 90° angle, 1/4 for 45°, and 0/4 for 0°. SCTS and radiography did not identify any of them.
Conclusion The results indicate that ECTS reconstructions are useful for assessment of metal screws and for detection of nondisplaced screw fractures.
Subject
General Veterinary,Animal Science and Zoology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献