Complications of Transcutaneous Protheses – A Systematic Review of Publications Over the Past 10 Years

Author:

Souza Marcos Antonio de1ORCID,Vallejos Riart Santiago Luis2ORCID,Souza Stephanie Rugeri de1,de Brito Rubens1ORCID,Bento Ricardo Ferreira1ORCID

Affiliation:

1. Department of Otorhinolaryngology, Universidade de São Paulo (USP), São Paulo, SP, Brazil

2. Department of Otorhinolaryngology, Universidad Nacional de Asunción (UNA), San Lorenzo, Central, Paraguay

Abstract

Abstract Introduction Bone conduction implants, responsible for transmitting sound from an external processor to the inner ear, can be divided into active and passive, depending on the vibratory stimulus location. The use of transcutaneous device has increased, given its aesthetic appeal, the complications and limitations of percutaneous devices, and patient's treatment adherence, focusing mainly on efficacy. However, various complications are associated with the use of transcutaneous prosthesis, which can often be serious. Objective To approach the literature on complications involving transcutaneous bone-anchored prostheses through a systematic review of articles published in the past 10 years (2011–2021). Data Synthesis The MEDLINE, EMBASE, Scopus, and Cochrane Library databases were searched. All articles written in English reporting on currently available transcutaneous prosthesis implantation and its complications were selected. Studies on both children and adults were included. The data on complications were extracted, and complications were classified as minor or major and associated to each device used. Thirty-seven articles were included in the study, of which 14 were prospective cohort studies, 22 were retrospective case series, and 1 was a case report. Most studies (18) included both adults and children. Moreover, 901 implantations were performed, of which 552 implanted Baha Attract (Cochlear Ltd., Sydney, Australia), 244 implanted BoneBridge (MED-EL, Innsbruck, Austria), and 105 implanted Sophono (Sophono Inc., Boulder, CO, USA]). Furthermore, 192 adverse events were reported (total complication rate, 21.3%), with 161 minor complications (84.3%) and 31 major complications (16.1%). Conclusion Transcutaneous prosthesis is an audiological alternative with fewer complications than percutaneous prosthesis. However, its indication should be judicious because complications are common, and although most complications are minor, serious infections requiring explantation may develop.

Publisher

Georg Thieme Verlag KG

Subject

Otorhinolaryngology

Reference32 articles.

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3