3D printing of fillable individual thyroid replicas based on nuclear medicine DICOM data used as phantoms for gamma probe calibration

Author:

Kühnel Christian1,Seifert Philipp1,Mulik Christina12,Winkens Thomas1,Freesmeyer Martin1

Affiliation:

1. Clinic of Nuclear Medicine, Jena-University-Hospital, Jena, Germany

2. Medical Engineering and Biotechnology University of Applied Sciences Jena (EAH-Jena), Jena, Germany

Abstract

Abstract Aim To proof the feasibility of manufacturing patient-indivdual (anthropomorphic) thyroid replicas from I-124 PET DICOM datasets by means of 3D printing. A possible field of application is the use of those phantoms for the calibration of gamma probes. Methods After editing of the DICOM datasets using several software types and transferring into a dedicated stereolithography format, 10 fillable thyroid replicas (35–200 mL) made of polylactide acid were manufactured via 3D printing. All replicas were filled with a water-solution containing 3.5 MBq I-131 and applied to a standard neck phantom. Calibration factor measurements were carried out using a clinical gamma probe. Measurements were performed with three different tilts: + 15°, 0° and –15°. The influence of the replicas’ volume and the tilt was investigated. Results Manufacturing of the replicas was successful in all cases. The time required for data processing was 13 ± 2 (median: 12, range: 9–25) min and 4–11 h for 3D printing (size-dependent). The printing process could be done overnight. Measured mean calibration factor for straight gamma probe positioning (0° tilt) was 31 965 ± 3360 (33 893, 25 470–34 253) cpm/MBq. A tilt of –15° resulted in lower calibration factors (–7.7 %), whereas a tilt of + 15° led to higher values (+ 9.5 %); p = 0.001. The calibration factors were highly inversely proportional correlated to the volume of the replicas (r = –0.91, p < 0.001). Conclusion 3D printing of patient-individual (anthropomorphic) fillable thyroid replicas was feasable for a large range of volumes. The study demonstrates the influence of the volume as well as the tilt of the measured object for calibration factor measurements with gamma probes.

Publisher

Georg Thieme Verlag KG

Subject

Radiology, Nuclear Medicine and imaging,General Medicine

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3