Comparability of C-Peptide Measurements – Current Status and Clinical Relevance

Author:

Hörber Sebastian123,Orth Matthias4,Fritsche Andreas235,Peter Andreas123

Affiliation:

1. Institute for Clinical Chemistry and Pathobiochemistry, Department for Diagnostic Laboratory Medicine, University Hospital Tübingen, Germany

2. Institute for Diabetes Research and Metabolic Diseases of the Helmholtz Center Munich at the University of Tübingen, Germany

3. German Center for Diabetes Research (DZD), München-Neuherberg, Germany

4. Institute of Laboratory Medicine, Vinzenz von Paul Kliniken gGmbH, Stuttgart, Germany

5. Department of Internal Medicine IV, Division of Endocrinology, Diabetology and Nephrology, University of Tübingen, Germany

Abstract

AbstractC-peptide is an increasingly used and established marker for beta cell function by assessing endogenous insulin secretion. Accurate and comparable C-peptide measurements are needed in clinical practice and research studies. For example, to calculate HOMA-indices, the C-peptide/glucose ratio, and the classification of recently published novel subgroups of diabetes and prediabetes have used C-peptide measurements. Although the process for standardization of C-peptide measurements is advanced, its full implementation is still missing; therefore, the current status of the comparability of C-peptide measurements using different immunoassays is unclear. Here we compared five widely used C-peptide immunoassays on different analyzers (Abbott ALINITY i, DiaSorin Liaison XL, Roche Cobas e411, Siemens Healthineers ADVIA Centaur XPT, and Immulite 2000 XPi) using serum samples covering the clinically relevant C-peptide concentration range. Although all investigated immunoassays are traceable to the international reference reagent for C-peptide (NIBSC code: 84/510), results of C-peptide measurements showed significant differences between analyzers in the entire concentration range, especially with increasing C-peptide concentrations. The mean bias was largest (36.6%) between results of the immunoassays by Roche and Siemens Healthineers (ADVIA Centaur XPT), and both assays revealed large discrepancies compared to immunoassays by Abbott, DiaSorin, and Siemens Healthineers (Immulite 2000 XPi). In contrast, the three latter assays showed similar C-peptide results (mean bias: 2.3% to 4.2%). Consequently, C-peptide discrepancies might affect clinical diagnosis and the interpretation of study results. Therefore, there is an urgent need to implement and finalize the standardization process of C-peptide measurements to improve patient care and the comparability of research studies.

Publisher

Georg Thieme Verlag KG

Subject

Endocrinology,General Medicine,Endocrinology, Diabetes and Metabolism,Internal Medicine

Reference22 articles.

1. The clinical utility of C-peptide measurement in the care of patients with diabetes;A G Jones;Diabet Med,2013

2. C-peptide is the appropriate outcome measure for type 1 diabetes clinical trials to preserve beta-cell function: Report of an ADA workshop, 21-22 October 2001;J P Palmer;Diabetes,2004

3. Pseudoinsulinoma in a white man with autoimmune hypoglycemia due to anti-insulin antibodies: value of the free C-Peptide assay;A E Arzamendi;Am J Clin Pathol,2014

4. C-peptide determination in the diagnosis of type of diabetes and its management: A clinical perspective;E Maddaloni;Diabetes Obes Metab,2022

5. Novel subgroups of adult-onset diabetes and their association with outcomes: A data-driven cluster analysis of six variables;E Ahlqvist;Lancet Diabetes Endocrinol,2018

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3