Abstract
Abstract
Because of discrepancies between electronically and manually measured values of mean cell hemoglobin concentration (MCHC) encountered in studies of pathologic red cells, we studied the effect of cell water content on MCHC measurements by both methods. A series of red cell samples with varying water contents (54%-164% normal) were prepared from normal cells using the antibiotic nystatin. MCHC was then measured, using the microhematocrit centrifuge and three different electronic cell counters in common laboratory use. For MCHC values above 36 g/dl as measured by the spun hematocrit method, all three electronic counters under estimmated the MCHC, with increasing error as the true MCHC increased. For MCHC values below 30 g/dl, the values from two conductivity based instruments agreed with those from the spun hematocrit method, whereas one instrument based on light scattering overestimated the MCHC. These results indicate that inaccuracies in the measured mean cell volume (MCV) of dehydrated or otherwise undeformable cells may lead to spurious values for MCHC when electronic cell counters are used.