A novel in-line high frequency interrupter for use with bubble CPAP: A feasibility study in a premature lamb model

Author:

Sivieri E.M.12,Eichenwald E.C.123,Abbasi S.123,Wolfson M.R.4

Affiliation:

1. Children’s Hospital of Philadelphia Newborn Care at Pennsylvania Hospital, Philadelphia, PA, USA

2. Division of Neonatology, Children’s Hospital of Philadelphia, Philadelphia, PA, USA

3. Department of Pediatrics, Perelman School of Medicine, University of Pennsylvania, Philadelphia, PA, USA

4. Department of Physiology, Department of Pediatrics, Department of Thoracic Medicine and Surgery, Temple Lung Center, Center for Translational Medicine, Lewis Katz School of Medicine, Temple University, Philadelphia, PA, USA

Abstract

BACKGROUND: Recent in vitro testing of high frequency (HF) oscillation applied to bubble continuous positive airway pressure (BCPAP) using a novel flow interrupter device (HFI) demonstrated significantly improved CO2 washout while not altering delivered mean airway pressure (MAP) in a premature infant lung model. This study’s aim was to evaluate the safety and efficacy of the HFI paired with BCPAP in an animal model of prematurity prior to clinical testing. DESIGN/METHODS: Twelve fetal lambs, 131–135 days gestation, weight 3.51±0.42 kg, were delivered by Cesarean section. The lambs were supported by mechanical ventilation and weaned to spontaneous breathing with BCPAP at 6 cmH2O. A combined CO2/airflow sensor measured end-tidal (EtCO2) and tidal volume (VT). Blood gases, heart rate (HR), arterial pressure (Part), minute ventilation (MV), MAP, ventilatory efficiency index (VEI), thoracoabdominal phase angle and labored breathing index (LBI) were recorded over a 10-minute baseline period followed by four randomized 10-minute intervals with HFI set to either 8, 10, 12 or 15 Hz. RESULTS: EtCO2 decreased from baseline by 11.1±2.2SE%, 16.6±4.3SE%, 13.5±4.9SE%, and 19.5±4.5SE% at 8, 10, 12, and 15 Hz respectively (p < 0.001). Blood gases, SpO2, HR, Part, MAP, VT, MV, esophageal pressure, phase angle, and LBI underwent no significant change with HF. Respiratory rate decreased, and VEI increased, by 14.9±4.5SD% (p = 0.037) and 83±22SD% (p < 0.011) respectively, averaged over all frequencies. CONCLUSIONS: We demonstrated the safety and efficacy of a novel BCPAP flow interrupter device. HF applied to the respiratory system resulted in significantly improved CO2 clearance and ventilation efficiency with no deleterious physiological effects in a pre-term lamb model.

Publisher

IOS Press

Subject

Pediatrics, Perinatology and Child Health

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3