Validating the accuracy of the Omron HBP-M4500 blood pressure measuring device according to the ISO81060-2:2018+ Amd 1:2020 protocol

Author:

Takahashi Hakuo1,Yamashita Shingo21,Yakura Nobuki21

Affiliation:

1. Department of Cardiology, Biwako Yoikuin Hospital, Shiga, Japan

2. Department of Technology Development, Omron Healthcare Co., Ltd., Kyoto, Japan

Abstract

BACKGROUND: Blood pressure (BP) naturally undergoes fluctuations and variations, particularly during anesthesia administration during surgery, hemodialysis, upper and lower gastrointestinal endoscopy, exercise testing, arterial and venous catheterization, and rehabilitation. These changes in BP may lead to life-threatening events. OBJECTIVES: The performance of the Omron HBP-M4500 device in monitoring blood pressure (BP) in the upper arm was validated according to the International Organization for Standardization (ISO) 81060-2:2018+amendment (Amd) 1:2020 protocol. METHODS: The device was used to assess 113 participants in the inflation mode, and 107 participants in the deflation mode. All the patients fulfilled the inclusion criteria, including the arm circumference range and systolic and diastolic BP levels, outlined in the protocol. Data validation and analysis were performed according to the manufacturer’s instructions. RESULTS: In criterion 1, the mean ± standard deviation (SD) values of the differences between the test device and reference BP were -0.6 ± 5.80/2.8 ± 6.78 mmHg (systolic/diastolic) and -1.0 ± 5.35/3.2 ± 6.52 mmHg for the inflation and deflation modes, respectively. These data fulfilled the ISO81060-2:2018+Amd1:2020 requirements of ⩽ 5 ±⁣⩽ 8 mmHg. In criterion 2, the differences were -0.6 ± 4.44/2.8 ± 6.26 and -1.0 ± 3.84/3.2 ± 6.09 mmHg for the inflation and deflation modes, respectively, fulfilling criterion 2 with SD values of ⩽ 6.91 and ⩽ 6.87 for systolic BP and ⩽ 6.34 and ⩽ 6.14 for diastolic BP in the inflation and deflation modes, respectively. These two criteria were fulfilled in both studies. CONCLUSION: The Omron HBP-M4500 device, either in inflation or deflation mode, fulfilled the criteria outlined in the ISO protocol. Therefore, this device is valuable for BP measurement in clinical and hospital settings.

Publisher

IOS Press

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3