INR calibration of Owren-type prothrombin time based on the relationship between PT% and INR utilizing normal plasma samples

Author:

Egberg Nils,Hillarp Andreas,Ødegaard Ole,Edlund Bror,Svensson Jan,Sandset Per,Rånby Mats,Lindahl Tomas

Abstract

SummaryProthrombin time (PT) is clinically important and is used to monitor oral anticoagulant therapy. To obtain PT results in international normalized ratio (INR), the current standardization procedure is complex and involves reference reagents.The PT of diluted plasma samples can be determined with a combined thromboplastin (the Owren-type procedure), but not necessarily with a plain thromboplastin (the Quick-type procedure). Owren-type PT procedures can therefore, as an alternative to the INR calibration, be calibrated with diluted normal plasma to give PT results in percent of normal PT activity (PT%).The present study explored if a plasma-based calibration of an Owren-type PT procedure can be used to obtain results in INR. The approach was to establish a relationship between PT% and INR by multi-center analysis of 365 samples from healthy individuals and patients on warfarin treatment. INR values were obtained by manual Quick-type reference procedure and PT% values by various automated Owren-type procedures. A relationship INR = (1/PT% + 0.018)/0.028 was found. A calibration procedure, based on the relationship, was investigated. Calibrators were the median PT of 21 normal plasma at dilutions representing 100%, 50%, 25%, 12.5% and 6.25% of normal PT activity. These were assigned INR values of 1.00, 1.36, 2.07, 3.05 and 6.36. Calibration of various Owren-type assays was repeatedly performed by 5 expert laboratories during 3 consecutive years. The INR values of certain lyophilised or frozen control plasmas were determined. The frozen control plasmas had externally assigned INR values according to WHO guidelines. Within the laboratory, CV was typically below 3%. No appreciable difference among the results of the different laboratories or the three assay occasions was found. Externally assigned and INR values were essentially identical to those found. These and other results indicated that the calibration procedure was reproducible, precise and accurate. Thus, an Owren-type PT assay can be calibrated with normal plasma samples to give results in INR and the investigated calibration procedure can be proposed for this purpose.

Publisher

Georg Thieme Verlag KG

Subject

Hematology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3