Intrarater and Interrater Reliability of Quantitative Ultrasound Speed of Sound by Trained Raters at the Distal Radius in Postmenopausal Women

Author:

Watson Cynthia J.1,de Ruig Matthew J.1,Saunders Kent T.2

Affiliation:

1. School of Physical Therapy, Anderson University, Greenville, South Carolina.

2. College of Business, Anderson University, Greenville, South Carolina.

Abstract

ABSTRACT Background and Purpose: Quantitative ultrasound (QUS) performed at the distal radius shows potential as an effective tool for evaluating fracture risk, especially in postmenopausal women, but the reliability of this tool has not been reported in this patient population. This study investigated the within-day intrarater and interrater reliability of trained raters performing QUS measures at the distal radius of postmenopausal community-dwelling women. Methods: Eighteen women between the ages of 56 and 87 years were recruited from senior centers in the Greenville area of South Carolina and consented to participate. A QUS bone sonometer was used in this study; the trained raters performed 3 repeated speed of sound (SOS) measures at the nondominant radius of each participant on the same day. Intraclass correlation coefficients (ICCs) were calculated (model 3,1 for intrarater and 2,3 for interrater reliability) to determine the reliability within and between raters. In addition, we calculated the standard error of measurement (SEM) and the minimal detectable change at 95% confidence interval (MDC95). Results: The within-day intrarater reliability was excellent for both physical therapist raters (ICC(3,1) ranging from 0.93 to 0.91). The SEM and MDC for the within-day intrarater reliability measures were small for both raters (SEM = 36.79-38.75 or 0.92%-0.97% of the pooled mean; MDC95 = 101.99-107.40). The interrater reliability for SOS measurements using radial QUS was good with an ICC(2,3) of 0.77 (95% CI, 0.36-0.92) for 3 averaged measures. The SEM and MDC were larger for the interrater reliability measures (SEM = 62.79 or 1.57% of the pooled mean; MDC95 = 174.05). Conclusions: Quantitative ultrasound SOS measurement at the distal radius showed excellent within-day intrarater and good interrater (when 3 measures were averaged) reliability in community-dwelling postmenopausal women. The SEM and MDC95 were also low for interrater reliability when a mean of 3 trials was used. This study is the first to demonstrate that this particular QUS device at the radius may be a reliable fracture risk screening tool for postmenopausal women.

Publisher

Ovid Technologies (Wolters Kluwer Health)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3