Abstract BACKGROUND Cerebrospinal fluid (CSF) amyloid-β (Aβ42) is a well-established biomarker for Alzheimer disease. Several immunoassays for Aβ42 exist but differ in absolute concentrations and may suffer from matrix interference, thereby hampering interlaboratory comparisons and the use of general cutoff levels. Together with the IFCC Working Group on CSF Proteins, we developed a candidate reference measurement procedure (RMP) for Aβ42. METHODS The antibody-independent candidate RMP was based on solid-phase extraction and isotope-dilution LC-MS/MS. The candidate RMP used 2 differently stable isotope-labeled Aβ42 peptides for calibration in human CSF, an important aspect since there was no analyte-free matrix available. Because no CSF certified reference material (CRM) exists, we used a nonlabeled Aβ42 standard, the concentration of which was determined by amino acid analysis. We performed measurements on a high-resolution quadrupole-Orbitrap hybrid instrument. The results were compared with a method run in a second laboratory with triple quadrupole instrumentation. RESULTS The candidate RMP allowed quantification of CSF Aβ42 from 150 to 4000 pg/mL. Validation of the method showed a recovery of 100% (15%), intraassay and interassay imprecision of 5.0% and 6.4%, respectively, and an expanded uncertainty of 15.7%. No analytical interferences or carryover were detected. CONCLUSIONS This method will help set the value of CSF Aβ42 in a CRM, which could be used to harmonize Aβ42 assays and facilitate the introduction of general cutoff concentrations for CSF Aβ42 in clinical trials and practice.

Funder

Demensförbundet

Magnus Bergvall's Foundation

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

Cited by 128 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3