Candidate Reference Measurement Procedure for the Quantification of Total Serum Cortisol with LC-MS/MS

Author:

Hawley James M1,Owen Laura J12,MacKenzie Finlay3,Mussell Chris4,Cowen Simon4,Keevil Brian G12

Affiliation:

1. University Hospital South Manchester, Manchester, UK

2. Manchester Academic Health Science Centre, Manchester, UK

3. Birmingham Quality, UK National External Quality Assessment Scheme, University Hospitals Birmingham National Health Service Foundation Trust, Birmingham, UK

4. LGC Ltd., Teddington, Middlesex, UK

Abstract

Abstract BACKGROUND Accurate measurement of serum cortisol is required to diagnose and treat adrenal disorders. Although certified reference materials (CRMs) are available to standardize cortisol measurements, External Quality Assessment (EQA) schemes still demonstrate a wide dispersion of results. We present a serum cortisol candidate reference measurement procedure that, through analysis of a Joint Committee for Traceability in Laboratory Medicine–listed panel of higher-order CRMs, provides metrologically traceable results. METHOD Isotope-labeled internal standard was added to samples before supported liquid extraction. Extracts were analyzed with LC-MS/MS in positive electrospray ionization mode. Multiple reaction monitoring was used to detect cortisol and its corresponding internal standard transitions. We measured samples in triplicate over 3 days and calculated the mean result. RESULTS Mean intra- and interassay imprecision were 1.3% and 1.5%, respectively, for concentrations of 154, 510, and 769 nmol/L. Ionization efficiency studies and structural analog analysis proved the method to be robust against interferences. Through analysis of 34 CRMs (83–764 nmol/L), expanded measurement uncertainty was calculated to be 5% (95% CI). The mean bias between the measured and target CRM concentrations was statistically insignificant at −0.08%. CONCLUSIONS The accuracy and low measurement uncertainty of this method qualify it as a CRM procedure. Metrological traceability has been achieved through the analysis of higher-order CRMs. This method could be used to underpin serum cortisol EQA schemes to provide samples with a traceable target value, enabling participating laboratories to determine the accuracy and measurement uncertainty of their assays.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3