Background: High-performance liquid chromatography–tandem mass spectrometric (LC-MS/MS)1 analysis of plasma free metanephrines is the most diagnostically sensitive and specific screening test for the diagnosis of pheochromocytoma. We sought to develop an in-house method for this expensive test
Methods: We used off-line isopropanol protein precipitation of plasma to remove interfering substances before LC-MS/MS analysis. We compared the extraction efficiency and limits of quantification of protein precipitation to those of previously reported solid-phase techniques.
Results: The new method had limits of quantification of 0.09 nmol/L and 0.17 nmol/L for metanephrine and normetanephrine, respectively. Method comparison with a previously described solid-phase extraction method revealed Deming regression slopes of 0.904 and 0.994, intercepts of 0.007 and 0.023, and SEs of the residuals (Sy|x) of 0.071 and 0.284 for metanephrine and normetanephrine, respectively. Extraction efficiency of isopropanol protein precipitation was 66% for metanephrine and 35% for normetanephrine, results that were superior to the efficiencies of 4% and 1% for our adapted solid-phase extraction method. No ion suppression was observed at the retention times for metanephrine and normetanephrine.
Conclusions: Isopropanol protein precipitation is a novel and effective off-line sample preparation method for metanephrines that offers a less expensive alternative to on-line solid-phase extraction for low-volume testing and requires a sample volume of only 200 μL. The mass spectrometric analysis time is equivalent to that of solid-phase techniques.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
44 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献