Simultaneous Measurement of Plasma Concentrations of Proinsulin and C-Peptide and Their Ratio with a Trefoil-Type Time-Resolved Fluorescence Immunoassay

Author:

De Pauw Pieter E M1,Vermeulen Ilse1,Ubani Ogonnaya C1,Truyen Inge1,Vekens Evilien M F1,van Genderen Farah T1,De Grijse Joeri W1,Pipeleers Daniel G1,Van Schravendijk Chris1,Gorus Frans K1

Affiliation:

1. Diabetes Research Center, Brussels Free University, Brussels, Belgium

Abstract

Abstract Background: When the concentrations of 2 or more substances are measured separately, their molar ratios are subject to the additive imprecisions of the different assays. We hypothesized that the cumulative error for concentration ratios of peptides containing a common sequence might be minimized by measuring the peptides simultaneously with a “trefoil-type” immunoassay. Methods: As a model of this approach, we developed a dual-label time-resolved fluorescence immunoassay (TRFIA) to simultaneously measure proinsulin, C-peptide, and the proinsulin–C-peptide ratio (PI/C). A monoclonal antibody captures all C-peptide–containing molecules, and 2 differently labeled antibodies distinguish between proinsulin-like molecules and true C-peptide. Results: The trefoil-type TRFIA was capable of measuring plasma C-peptide and proinsulin simultaneously without mutual interference at limits of quantification of 48 and 8125 pmol/L, and 2.1 and 197 pmol/L, respectively. Within-laboratory imprecision values for the trefoil-type TRFIA ranged between 8.4% and 12% for the hormone concentrations. Unlike the hormone results obtained with separate assays, imprecision did not increase when PI/C was calculated from trefoil assay results (P < 0.05). Peptide concentrations were highly correlated with results obtained in individual comparison assays (r2 ≥ 0.965; P < 0.0001). The total error for PI/C obtained with the trefoil-type TRFIA remained ≤25% over a broader C-peptide range than with separate hormone assays (79–7200 pmol/L vs 590–4300 pmol/L C-peptide). Preliminary data indicate little or no interference by heterophile antibodies. Conclusions: The developed trefoil-type TRFIA is a reliable method for simultaneous measurement of proinsulin, C-peptide, and PI/C and provides proof of principle for the development of other trefoil-type multiple-label immunoassays.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

Cited by 26 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3